DXA scanning of acetabulum in patients with cementless total hip arthroplasty.

作者信息

Laursen Mogens B, Nielsen Poul T, Søballe Kjeld

机构信息

Northern Orthopedic Division, Aalborg University Hospital, part of Aarhus University, Aalborg, Denmark.

出版信息

J Clin Densitom. 2005 Winter;8(4):476-83. doi: 10.1385/jcd:8:4:476.

Abstract

The aim of this study was to evaluate the reproducibility of bone mineral density (BMD) measurements of the periprosthetic bone in patients with hemispherical acetabular cups in cementless total hip arthroplasty (THA). Thirty patients were treated for primary osteoarthrosis with cementless THA. Dual-energy X-ray absorptiometry (DXA) scanning was performed with a pencil-beam bone densitometer (Norland XR-36). Accuracy and reproducibility was determined by double measurements of BMD in four regions of interest (ROI). The influence of patient postures including various pelvic inclination angles was evaluated as well. Pitman test for a combined netROI revealed a standard deviation ratio of 3.2 for the anterio-posterior scans related to the lateral position. The Wilkinson ROIs showed a high intraobserver agreement. With pelvic tilt increasing until 20 degrees , the precision of DXA scanning decreased. In conclusion, reproducibility of DXA scanning was high. This study demonstrated that the patients can be scanned in the supine position, and BMD measurement of the periacetabular bone can be performed using the Wilkinson model with four rectangular ROIs.

摘要

本研究的目的是评估非骨水泥型全髋关节置换术(THA)中使用半球形髋臼杯的患者假体周围骨骨密度(BMD)测量的可重复性。30例原发性骨关节炎患者接受了非骨水泥型THA治疗。使用笔形束骨密度仪(Norland XR - 36)进行双能X线吸收法(DXA)扫描。通过在四个感兴趣区域(ROI)对BMD进行两次测量来确定准确性和可重复性。还评估了包括各种骨盆倾斜角度在内的患者姿势的影响。对组合净ROI的皮特曼检验显示,与侧位相关的前后位扫描的标准差比为3.2。威尔金森ROI显示出较高的观察者内一致性。随着骨盆倾斜增加至20度,DXA扫描的精度降低。总之,DXA扫描的可重复性很高。本研究表明,患者可以在仰卧位进行扫描,并且可以使用具有四个矩形ROI的威尔金森模型进行髋臼周围骨的BMD测量。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验