Vegfors M, Lindberg L G, Oberg P A, Lennmarken C
Department of Anaesthesiology, University Hospital, Sweden.
Acta Anaesthesiol Scand. 1992 Jul;36(5):454-9. doi: 10.1111/j.1399-6576.1992.tb03496.x.
The object of this study was to investigate the influence of haematocrit on the accuracy of pulse oximetry. Seven Swedish land race rabbits were studied. The oxygen saturation of haemoglobin was decreased step-wise using increasing fractions of nitrogen to the inspiratory gas. One pulse oximeter probe was attached on the front leg and another probe directly over the common carotid artery. The pulse oximeter readings (SpO2) were compared with simultaneous oxygen saturation analysis (SaO2) by a haemoximeter. The pulse oximeter measurements were performed at the haematocrit levels of approximately 40% and 11%, respectively. We found a good correlation between SpO2 and SaO2 in a wide range of the oxygen saturation, i.e. SaO2 26-100%. After haemodilution the correlation was improved in the range 86-100%, but not in the range 26-85%. No correlation between SpO2 and SaO2 was found when the sensor was attached directly over the artery during normal haematocrit levels. After haemodilution a better correlation was however obtained. These results indicate that the accuracy of pulse oximetry is dependent on the haematocrit level.
本研究的目的是调查血细胞比容对脉搏血氧饱和度测定准确性的影响。对7只瑞典本地品种兔子进行了研究。通过向吸入气体中增加氮气比例,逐步降低血红蛋白的氧饱和度。在前腿上连接一个脉搏血氧仪探头,在颈总动脉正上方连接另一个探头。将脉搏血氧仪读数(SpO2)与通过血液氧含量计同时进行的血氧饱和度分析(SaO2)进行比较。脉搏血氧仪测量分别在血细胞比容水平约为40%和11%时进行。我们发现在较宽的氧饱和度范围内,即SaO2为26% - 100%时,SpO2与SaO2之间存在良好的相关性。血液稀释后,在86% - 100%的范围内相关性得到改善,但在26% - 85%的范围内没有改善。当传感器在正常血细胞比容水平时直接连接在动脉上方时,未发现SpO2与SaO2之间存在相关性。然而,血液稀释后获得了更好的相关性。这些结果表明脉搏血氧饱和度测定的准确性取决于血细胞比容水平。