Good Gregory W, Schepler Adrianne, Nichols Jason J
The Ohio State University College of Optometry, Columbus, 43210, USA.
Optom Vis Sci. 2005 Dec;82(12):1054-9. doi: 10.1097/01.opx.0000192351.63069.4a.
The Lanthony Desaturated Panel D-15 has been used to measure fine color discrimination for congenital and acquired color vision defects. This study investigated the test-retest reliability of the test using an intertest interval of approximately 1 month.
One hundred twenty-six color vision normals (mean age = 34.5 years) were administered several color vision tests, including the Lanthony Desaturated D-15. Normal color vision status was confirmed using the anomaloscope and HRR color plates. The color vision tests were readministered 3 to 6 weeks after initial testing. The results of the Lanthony test were expressed using the color confusion index of Bowman. The difference in Color Confusion Index (CCI) between the two administrations was calculated and used to determine the intraclass correlation coefficient.
The overall mean CCI for the two administrations for these subjects was 1.11 +/- 0.136. The mean difference in CCI score between test administrations was -0.02 +/- 0.128. There was a strong correlation between the absolute value of the difference in CCI and the mean CCI for each subject (r = 0.51, p < 0.0001). The intraclass correlation coefficient was 0.56 (95% confidence interval, 0.43-0.67).
Although the Lanthony Desaturated D-15 test can be used to assess fine color discrimination, there is considerable within-subject variability in test results. The intraclass correlation coefficient is less than that recommended for use in clinical testing or research. Clinicians should consider at least three administrations of the test at each sitting to ensure precision and we recommend taking the mean of those three tests.
兰托尼去饱和 Panel D - 15 已被用于测量先天性和后天性色觉缺陷的精细颜色辨别能力。本研究使用约 1 个月的测试间隔来调查该测试的重测信度。
对 126 名色觉正常者(平均年龄 = 34.5 岁)进行了多项色觉测试,包括兰托尼去饱和 D - 15 测试。使用检眼镜和 HRR 色盲本确认正常色觉状态。在初次测试后 3 至 6 周重新进行色觉测试。兰托尼测试的结果用鲍曼颜色混淆指数表示。计算两次测试之间颜色混淆指数(CCI)的差异,并用于确定组内相关系数。
这些受试者两次测试的总体平均 CCI 为 1.11±0.136。两次测试之间 CCI 得分的平均差异为 -0.02±0.128。每个受试者的 CCI 差异绝对值与平均 CCI 之间存在强相关性(r = 0.51,p < 0.0001)。组内相关系数为 0.56(95%置信区间,0.43 - 0.67)。
尽管兰托尼去饱和 D - 15 测试可用于评估精细颜色辨别能力,但测试结果在受试者内部存在相当大的变异性。组内相关系数低于临床测试或研究推荐使用的值。临床医生每次就诊时应至少进行三次该测试以确保准确性,并且我们建议取这三次测试的平均值。