一种适用于各种二氧化碳(CO₂)应用的多功能双通道二氧化碳(CO₂)注入器。原型。

A versatile dual-channel carbon dioxide (CO2) insufflator for various CO2)applications. The prototype.

作者信息

Nakajima K, Yasumasa K, Endo S, Takahashi T, Nishitani A, Nezu R, Nishida T

机构信息

Department of Surgery E1, Osaka University Graduate School of Medicine, 2-2, Yamadaoka, Suita Osaka, 565-0871, Japan.

出版信息

Surg Endosc. 2006 Feb;20(2):334-8. doi: 10.1007/s00464-005-0208-4. Epub 2005 Dec 9.

Abstract

BACKGROUND

Carbon dioxide (CO2), with its rapid absorptive nature, has been proven superior to atmospheric air as an insufflating agent in various clinical settings. However, CO2 insufflation has not gained wide clinical acceptance, mainly because there has been no suitable feeding system. The authors therefore have developed a versatile "dual-channel" CO2 insufflator that facilitates wider use of CO2. The objectives of this study were to introduce the authors' prototype insufflator, to evaluate its safety and performance, and to validate CO2 application using the prototype.

METHODS

The prototype insufflator provides one CO2 inlet connected to a regular CO2 gas cylinder and two CO2 outlets positioned on the front and back of the device, respectively. The CO2 gas fed from the cylinder is pressure-regulated and divided into two independent conduits inside the device. The front outlet feeds CO2 gas for pneumoperitoneum at an electronically controlled pressure and flow rate. The back channel supplies CO2 gas at a fixed flow rate, allowing manual control of insufflation for various purposes. The device was evaluated with canine models.

RESULTS

The prototype was safe and performed well. The CO2 application (colonoscopy in this series) using the back channel was feasible while intact CO2 pneumoperitoneum was simultaneously maintained via the front channel. There were no device malfunctions. The serial abdominal x-rays indicated that intraluminal CO2 insufflation such as that used for CO2 colonoscopy caused less residual intestinal gas than conventional air insufflation.

CONCLUSIONS

The dual-channel CO2 insufflator enabled two different modes of CO2 insufflation at the same time from a single CO2 cylinder. The authors are now improving the prototype to allow safer and wider usage of CO2 in the operating room.

摘要

背景

二氧化碳(CO₂)具有快速吸收的特性,在各种临床环境中,已被证明作为一种注入剂优于大气空气。然而,CO₂注入尚未得到广泛的临床认可,主要是因为没有合适的输送系统。因此,作者开发了一种多功能的“双通道”CO₂注入器,以促进CO₂的更广泛应用。本研究的目的是介绍作者的原型注入器,评估其安全性和性能,并使用该原型验证CO₂的应用。

方法

原型注入器有一个连接到常规CO₂气瓶的CO₂入口,以及分别位于设备前后的两个CO₂出口。从气瓶输送来的CO₂气体在设备内部进行压力调节,并分成两个独立的管道。前出口以电子控制的压力和流速输送用于气腹的CO₂气体。后通道以固定流速供应CO₂气体,允许手动控制用于各种目的的注入。该设备在犬模型上进行了评估。

结果

该原型安全且性能良好。使用后通道进行CO₂应用(本系列中为结肠镜检查)是可行的,同时通过前通道可同时维持完整的CO₂气腹。没有设备故障。系列腹部X线检查表明,用于CO₂结肠镜检查的腔内CO₂注入比传统空气注入产生的肠道残留气体更少。

结论

双通道CO₂注入器能够从单个CO₂气瓶同时实现两种不同模式的CO₂注入。作者目前正在改进该原型,以在手术室中更安全、更广泛地使用CO₂。