Rimoldi Ornella, Schäfers Klaus P, Boellaard Ronald, Turkheimer Federico, Stegger Lars, Law Marilyn P, Lammerstma Adriaan A, Camici Paolo G
Medical Research Council Clinical Sciences Centre, Faculty of Medicine, Imperial College of Science, Technology and Medicine, Hammersmith Hospital, London, United Kingdom.
J Nucl Med. 2006 Jan;47(1):163-72.
The purpose of this study was to assess the feasibility and accuracy of quantifying subendocardial and subepicardial myocardial blood flow (MBF) and the relative coronary flow reserves (CFR) using (15)O-labeled water (H(2)(15)O) and 3-dimensional-only PET.
Eight pigs were scanned with H(2)(15)O and (15)O-labeled carbon monoxide (C(15)O) after partially occluding the circumflex (n = 3) or the left anterior descending (n = 5) coronary artery, both at rest and during hyperemia induced by intravenous dipyridamole. Radioactive microspheres were injected during each of the H(2)(15)O scans.
In a total of 256 paired measurements of MBF, ranging from 0.30 to 4.46 mL.g(-1).min(-1), microsphere and PET MBF were fairly well correlated. The mean difference between the 2 methods was -0.01 +/- 0.52 mL.g(-1).min(-1) with 95% of the differences lying between the limits of agreement of -1.02 and 1.01 mL.g(-1).min(-1). CFR was significantly reduced (P < 0.05) in the ischemic subendocardium (PET = 1.12 +/- 0.45; microspheres = 1.09 +/- 0.50; P = 0.86) and subepicardium (PET = 1.2 +/- 0.35; microspheres = 1.32 +/- 0.5; P = 0.39) in comparison with remote subendocardium (PET = 1.7 +/- 0.62; microspheres = 1.64 +/- 0.61; P = 0.68) and subepicardium (PET = 1.79 +/- 0.73; microspheres = 2.19 +/- 0.86; P = 0.06).
Dynamic measurements using H(2)(15)O and a 3-dimensional-only PET tomograph allow regional estimates of the transmural distribution of MBF over a wide flow range, although transmural flow differences were underestimated because of the partial-volume effect. PET subendocardial and subepicardial CFR were in good agreement with the microsphere values.
本研究的目的是评估使用(15)O标记水(H₂(15)O)和仅三维PET定量心内膜下和心外膜下心肌血流量(MBF)以及相对冠状动脉血流储备(CFR)的可行性和准确性。
8头猪在部分闭塞回旋支(n = 3)或左前降支(n = 5)冠状动脉后,于静息状态和静脉注射双嘧达莫诱导的充血状态下,用H₂(15)O和(15)O标记的一氧化碳(C(15)O)进行扫描。在每次H₂(15)O扫描期间注射放射性微球。
在总共256对MBF测量值中,范围为0.30至4.46 mL·g⁻¹·min⁻¹,微球法和PET法测量的MBF相关性良好。两种方法之间的平均差异为-0.01±0.52 mL·g⁻¹·min⁻¹,95%的差异落在-1.02至1.01 mL·g⁻¹·min⁻¹的一致性界限内。与远心内膜下(PET = 1.7±0.62;微球法 = 1.64±0.61;P = 0.68)和心外膜下(PET = 1.79±0.