预浸泡及在超声波清洗器中的时长对根管治疗器械消毒后清洁度的影响

The effect of pre-soaking and time in the ultrasonic cleaner on the cleanliness of sterilized endodontic files.

作者信息

Aasim S A, Mellor A C, Qualtrough A J E

机构信息

Department of Restorative Dentistry, Dundee Dental School, Dundee, UK.

出版信息

Int Endod J. 2006 Feb;39(2):143-9. doi: 10.1111/j.1365-2591.2006.01058.x.

Abstract

AIMS

To assess whether pre-soaking files in an enzymatic cleaner prior to ultrasonic cleaning had any effect on cleanliness and also to assess the effect of the time that endodontic files spend in an ultrasonic bath prior to sterilization on their overall cleanliness.

METHODOLOGY

Twenty root canals in a total of ten patients were cleaned and shaped using conventional techniques. Following use, some of the files were pre-soaked and then ultrasonically cleaned for either 5, 10, 30 or 60 min. Other files had no pre-soaking and were then ultrasonically cleaned. There were two control groups, one where the files were pre-soaked and not ultrasonically cleaned and the other where the files were neither pre-soaked nor ultrasonically cleaned. All files were then subjected to a standard packing and autoclaving process. Following autoclaving, the files were examined using a light microscope at a magnification of 40x. The cutting section of each file was divided into two parts, the tip and the shaft, for visualization under the microscope. Any debris or cement on the files was scored using a modification of the scale used by Smith et al. (Journal of Hospital Infection, 51, 2002, 233). The data were analysed using one-way analysis of variance.

RESULTS

Pre-soaking had no significant effect on the cleanliness of the files (P = 0.18 at the tip, P = 0.93 at the shaft). Ultrasonic cleaning had a significant effect on the cleanliness of the files (P < 0.00) but there was not a linear relationship between cleanliness and the ultrasonic cleaning time. There was little benefit in extending the ultrasonic cleaning time beyond 5 min. Calcium hydroxide deposits on two files were resistant to ultrasonic cleaning.

CONCLUSIONS

There is no benefit in pre-soaking endodontic files prior to ultrasonic cleaning. The optimum time for ultrasonic cleaning was between 5 and 10 min. Further ultrasonic exposure, up to 60 min, did not improve cleanliness. Although a majority of files were free from debris following ultrasonic cleaning, a substantial minority still retained debris. This supports the case for endodontic files being single-use only.

摘要

目的

评估在超声清洗前将锉针浸泡于酶清洁剂中是否会对清洁度产生影响,同时评估根管锉针在灭菌前于超声清洗槽中浸泡的时间对其整体清洁度的影响。

方法

采用传统技术对总共10例患者的20个根管进行清理和塑形。使用后,部分锉针先进行预浸泡,然后超声清洗5、10、30或60分钟。其他锉针不进行预浸泡,直接进行超声清洗。设置两个对照组,一组锉针进行预浸泡但不进行超声清洗,另一组锉针既不进行预浸泡也不进行超声清洗。然后对所有锉针进行标准包装和高压灭菌处理。高压灭菌后,使用40倍放大倍数的光学显微镜检查锉针。将每个锉针的切割部分分为尖端和杆部两部分,以便在显微镜下观察。使用对Smith等人(《医院感染杂志》,51卷,2002年,233页)所使用的量表进行修改后的量表,对锉针上的任何碎屑或粘固剂进行评分。使用单因素方差分析对数据进行分析。

结果

预浸泡对锉针的清洁度无显著影响(尖端处P = 0.18,杆部处P = 0.93)。超声清洗对锉针的清洁度有显著影响(P < 0.00),但清洁度与超声清洗时间之间不存在线性关系。将超声清洗时间延长至5分钟以上几乎没有益处。两根锉针上的氢氧化钙沉积物对超声清洗有抗性。

结论

在超声清洗前预浸泡根管锉针没有益处。超声清洗的最佳时间为5至10分钟。进一步延长超声暴露时间至60分钟并不能提高清洁度。尽管大多数锉针在超声清洗后无碎屑残留,但仍有相当一部分锉针仍残留碎屑。这支持了根管锉针仅一次性使用的观点。