Suppr超能文献

无功能型艾哈迈德青光眼引流阀植入术的术中管理

Intraoperative management of nonfunctioning Ahmed glaucoma valve implant.

作者信息

Setabutr Pete, Bell Nicholas P F, Feldman Robert M

机构信息

Department of Ophthalmology and Visual Science, The University of Texas Medical School at Houston, Houston, Texas 77030, USA.

出版信息

Ophthalmic Surg Lasers Imaging. 2006 Jan-Feb;37(1):62-4.

Abstract

A 62-year-old woman with neovascular glaucoma underwent filtering surgery with the Ahmed glaucoma valve (New World Medical, Inc., Rancho Cucamonga, CA). Intraoperative hypotony secondary to a defective valve was noted and managed by excising the valve mechanism. The implant was then treated as if it was non-valved by placement of an obturator suture. The postoperative course was unremarkable for a non-valved implant. This case demonstrates how intraoperative hypotony secondary to overfiltration of an Ahmed glaucoma valve implant can be successfully managed by converting to a non-valved implant.

摘要

一名62岁的新生血管性青光眼女性患者接受了使用艾哈迈德青光眼引流阀(美国加利福尼亚州兰乔库卡蒙加市新世界医疗公司)的滤过手术。术中发现因引流阀缺陷导致的低眼压,并通过切除引流阀装置进行处理。然后通过放置封闭缝线将植入物当作无阀植入物来处理。术后过程对于无阀植入物而言并无异常。该病例表明,通过将艾哈迈德青光眼引流阀植入物转换为无阀植入物,可以成功处理因该植入物过度滤过导致的术中低眼压情况。