Suenami Koichi, Lim Lee Wah, Takeuchi Toyohide, Sasajima Yasuhide, Sato Kiyohito, Takekoshi Yuji, Kanno Susumu
Department of Chemistry, Faculty of Engineering, Gifu University, 1-1 Yanagido, Gifu, 501-1193, Japan.
Anal Bioanal Chem. 2006 Apr;384(7-8):1501-5. doi: 10.1007/s00216-006-0352-2. Epub 2006 Mar 14.
A rapid, selective and sensitive analytical method for the simultaneous determination of 16 nonsteroidal anti-inflammatory drugs (NSAIDs) in human plasma was carried out using Oasis HLB solid-phase extraction (SPE), followed by reversed-phase high-performance liquid chromatography and quadrupole mass spectrometry with electrospray ionization operated in the negative ion mode. The recoveries of NSAIDs from human plasma by SPE were greater than 76.7%. The use of a short column packed with small particles enabled rapid and simultaneous determination within 7 min. The detection limits for the NSAIDs were 0.01-0.9 mug/ml using the selected ion monitoring mode.