两款超声乳化仪相关热特性的比较。

Comparison of thermal features associated with 2 phacoemulsification machines.

作者信息

Brinton Jason P, Adams Wesley, Kumar Rajiv, Olson Randall J

机构信息

Department of Ophthalmology and Visual Sciences, University of Utah Health Sciences Center, Salt Lake City, Utah 84132, USA.

出版信息

J Cataract Refract Surg. 2006 Feb;32(2):288-93. doi: 10.1016/j.jcrs.2005.12.107.

Abstract

PURPOSE

To determine the thermal characteristics of the Legacy Advantec and Sovereign WhiteStar phacoemulsification machines during different clinically relevant scenarios.

SETTING

In vitro study.

METHODS

In water, temperature was recorded continuously on the sleeve in an artificial chamber, and the increase in temperature over baseline after 60 seconds of ultrasound was determined. This was done for continuous ultrasound, 50 ms on and 50 ms off (pulse), 6 ms on and 12 ms off (WhiteStar; Sovereign only) with aspiration blocked and not blocked, and with 100 g and 200 g weights suspended from the sleeve.

RESULTS

Comparing temperature increase per 20% machine power increments, Sovereign ran hotter than Legacy Advantec for continuous ultrasound (2.31x) and pulse (2.23x). Blocking aspiration increased temperature over the unblocked state. Pulsing decreased temperature by 51% (Legacy Advantec, pulse), 52% (Sovereign, pulse), and 64% (WhiteStar). Weights had much more effect on the Legacy Advantec: 3.5 times more going from baseline to 100 g weights and 3.2 times more going from 100 to 200 g weights. For all these comparisons, the P value was less than 0.0001.

CONCLUSIONS

The machines behaved fundamentally differently, with the Legacy Advantec controlling stroke length and Sovereign controlling a fixed power at any setting. Therefore, workload had a much bigger impact on Legacy Advantec thermal characteristics. Pulsing decreased heat produced directly related to the duty cycle. The most dangerous incision burn scenario is with continuous ultrasound, aspiration blocked, and a heavy workload.

摘要

目的

确定Legacy Advantec和Sovereign WhiteStar超声乳化仪在不同临床相关场景下的热特性。

设置

体外研究。

方法

在水中,在人工腔室的套管上连续记录温度,并确定超声60秒后相对于基线的温度升高。对连续超声、50毫秒开启和50毫秒关闭(脉冲)、6毫秒开启和12毫秒关闭(WhiteStar;仅适用于Sovereign)进行此操作,同时阻塞和不阻塞抽吸,并在套管上悬挂100克和200克的重物。

结果

比较每20%机器功率增量的温度升高,对于连续超声(2.31倍)和脉冲(2.23倍),Sovereign比Legacy Advantec运行时温度更高。阻塞抽吸会使温度高于未阻塞状态。脉冲使温度降低了51%(Legacy Advantec,脉冲)、52%(Sovereign,脉冲)和64%(WhiteStar)。重物对Legacy Advantec的影响更大:从基线到100克重物时温度升高3.5倍,从100克到200克重物时温度升高3.2倍。对于所有这些比较,P值均小于0.0001。

结论

这两款仪器的表现存在根本差异,Legacy Advantec控制冲程长度,而Sovereign在任何设置下控制固定功率。因此,工作量对Legacy Advantec的热特性影响更大。脉冲减少了与占空比直接相关的热量产生。最危险的切口烧伤情况是连续超声、抽吸阻塞且工作量较大时。