传统AdvanTec、千禧和至尊白星超声乳化系统的空气中热成像比较

In-air thermal imaging comparison of Legacy AdvanTec, Millennium, and Sovereign WhiteStar phacoemulsification systems.

作者信息

Olson Michael D, Miller Kevin M

机构信息

Jules Stein Eye Institute and the Department of Ophthalmology, David Geffen School of Medicine at UCLA, Los Angeles, California 90095-7002, USA.

出版信息

J Cataract Refract Surg. 2005 Aug;31(8):1640-7. doi: 10.1016/j.jcrs.2005.01.016.

Abstract

PURPOSE

To compare the temperature profiles of 3 popular phacoemulsification units (Alcon Legacy AdvanTec, Bausch & Lomb Millennium, and AMO Sovereign WhiteStar) under similar operating conditions in air.

SETTING

Jules Stein Eye Institute and the Department of Ophthalmology, David Geffen School of Medicine at UCLA, Los Angeles, California, USA.

METHODS

Phacoemulsification probes from the 3 units were placed side by side in air and imaged in the infrared region using model P60 ThermaCAM (Flir Systems). The highest temperature produced by each probe was measured 10 seconds and 30 seconds after power application. Testing was performed under conditions that might produce a corneal burn during cataract surgery. Irrigation flow was set at the low rate of 1 cc/min to simulate a tight incision. Aspiration flow was set at 0 cc/min to simulate occlusion of the needle lumen. Wound compression was simulated in some tests by suspending 22.6 g weights by rubber bands from the silicone sleeves. Manufacturers' specific and identical silicone sleeves were used to evaluate possible variations in thermal conductivity. The AdvanTec Legacy and Millennium were operated in pulse mode at 15 Hertz; 50% duty cycle; and 10%, 30%, and 50% power. The Sovereign WhiteStar was operated in both C/F (56 Hz, 33% duty cycle) and C/L (33 Hz, 20% duty cycle) modes at the same console power settings. Temperature profiles were determined at a variety of power settings with each system operating in continuous and pulse mode.

RESULTS

Under all experimental conditions (at 10%, 30%, and 50% powers; with and without external weights suspended from the phacoemulsification probes; with manufacturers' and identical silicone sleeves; and in continuous and pulse modes), the Millennium and the Sovereign WhiteStar generated higher temperatures than the Legacy AdvanTec.

CONCLUSIONS

Under controlled operating conditions in air and under a variety of power, load, and duty-cycle settings, the Millennium and the Sovereign WhiteStar, operating in both pulse and continuous modes, generated higher peak temperatures than the Legacy AdvanTec.

摘要

目的

比较3种常用超声乳化仪(爱尔康Legacy AdvanTec、博士伦Millennium和AMO Sovereign WhiteStar)在空气中相似操作条件下的温度分布情况。

设置

美国加利福尼亚州洛杉矶市加州大学洛杉矶分校大卫·格芬医学院朱尔斯·斯坦眼科研究所及眼科系。

方法

将3种仪器的超声乳化探头并排放置于空气中,使用P60 ThermaCAM(菲力尔系统公司)在红外区域进行成像。在施加能量后10秒和30秒测量每个探头产生的最高温度。测试在白内障手术中可能导致角膜烧伤的条件下进行。冲洗流量设定为低流量1毫升/分钟以模拟紧密切口。抽吸流量设定为0毫升/分钟以模拟针腔堵塞。在一些测试中,通过用橡皮筋将22.6克重物悬挂在硅胶套管上来模拟伤口压迫。使用制造商特定且相同的硅胶套管来评估热导率的可能变化。Legacy AdvanTec和Millennium以15赫兹的脉冲模式运行;占空比为50%;功率分别为10%、30%和50%。Sovereign WhiteStar在同一控制台功率设置下以C/F(56赫兹,33%占空比)和C/L(33赫兹,20%占空比)模式运行。在每个系统以连续和脉冲模式运行的各种功率设置下确定温度分布。

结果

在所有实验条件下(功率为10%、30%和50%;超声乳化探头悬挂重物与否;使用制造商的和相同的硅胶套管;以及连续和脉冲模式),Millennium和Sovereign WhiteStar产生的温度高于Legacy AdvanTec。

结论

在空气中受控的操作条件下以及在各种功率、负载和占空比设置下,以脉冲和连续模式运行的Millennium和Sovereign WhiteStar产生的峰值温度高于Legacy AdvanTec。