Abbiati G, Fossati T, Arrigoni M, Rolle P, Dognini G L, Castiglioni C
Biopharmaceutical Department, LPB Research Institute, Cinisello Balsamo, Italy.
J Chromatogr. 1991 May 31;566(2):445-51. doi: 10.1016/0378-4347(91)80261-a.
A high-performance liquid chromatographic method using an enzymic reactor for determination of L-alpha-glycerophosphorylcholine in pharmaceutical forms is described. The procedure includes incubation of L-alpha-glycerophosphorylcholine with glycerophosphorylcholine phosphodiesterase (EC 3.1.4.2), giving choline and glycerophosphate, and subsequent chromatography of choline with a post-column enzymic reactor and electrochemical detection. The results obtained show a close linearity of the whole assay from 2 to 150 nmol/ml L-alpha-glycerophosphorylcholine, the sensitivity being 2 pmol per 20 microliters of injected sample. The precision of the method in the analysis of L-alpha-glycerophosphorylcholine in pharmaceutical forms, ampoules and capsules, was 1.34 and 1.21%, respectively.