Suppr超能文献

羟基磷灰石双涂层和未涂层的钛多轴椎弓根螺钉:一种体内牛模型

Hydroxyapatite dip coated and uncoated titanium poly-axial pedicle screws: an in vivo bovine model.

作者信息

Yildirim Omer S, Aksakal Bunyamin, Hanyaloglu Salih C, Erdogan Fazli, Okur Ali

机构信息

Faculty of Medicine, Department of Orthopaedic Surgery, Ataturk University, Erzurum, Turkey.

出版信息

Spine (Phila Pa 1976). 2006 Apr 15;31(8):E215-20. doi: 10.1097/01.brs.0000210221.00778.c7.

Abstract

STUDY DESIGN

A 1-year-old calf was implanted with titanium pedicle screws either uncoated (n = 7) or coated with hydroxyapatite (n = 7) by the dipping method on the pedicles of vertebrae (t10-l3).

OBJECTIVE

To evaluate biomechanical and histomorphologic responses to titanium pedicle screws uncoated and coated with hydroxyapatite.

SUMMARY OF BACKGROUND DATA

Failure of fixation caused by loosening of pedicle screws is a problem in spinal surgery. Enhancement of the fixation ability of screws by coating with biocompatible materials may improve prognosis of surgery.

METHODS

The calf was euthanized 4 months after implantation for determination of insertion and extraction torques for screws, and histologic and scanning electron microscopic examinations of areas screw embedded.

RESULTS

Insertion torques did not differ by the kind of screws (99 +/- 5.7 Ncm). However, extraction torques for screws coated with hydroxyapatite were higher than for screws uncoated (249 vs. 133 Ncm, P < 0.01). As compared with uncoated screws, implanted areas for hydroxyapatite-coated screws were denser, had more cracks, and provided better bonding. Fibrous tissue and new bone formation were observed around the areas of uncoated and hydroxyapatite-coated screws embedded, respectively.

CONCLUSIONS

Hydroxyapatite coating of titanium pedicle screws by the dipping method improved fixation and vertebral bone-implant interface, suggesting a decreased risk of a screw-loosening problem.

摘要

研究设计

将一枚1岁小牛的椎弓根(T10-L3)植入未涂层(n = 7)或通过浸渍法涂有羟基磷灰石(n = 7)的钛椎弓根螺钉。

目的

评估未涂层和涂有羟基磷灰石的钛椎弓根螺钉的生物力学和组织形态学反应。

背景数据总结

椎弓根螺钉松动导致的固定失败是脊柱手术中的一个问题。通过用生物相容性材料涂层来增强螺钉的固定能力可能会改善手术预后。

方法

植入后4个月对小牛实施安乐死,以测定螺钉的插入和拔出扭矩,并对螺钉植入区域进行组织学和扫描电子显微镜检查。

结果

螺钉类型对插入扭矩没有影响(99±5.7 Ncm)。然而,涂有羟基磷灰石的螺钉的拔出扭矩高于未涂层的螺钉(249对133 Ncm,P<0.01)。与未涂层螺钉相比,涂有羟基磷灰石的螺钉植入区域更致密,有更多裂纹,且结合更好。分别在未涂层和涂有羟基磷灰石的螺钉植入区域周围观察到纤维组织和新骨形成。

结论

通过浸渍法对钛椎弓根螺钉进行羟基磷灰石涂层可改善固定和椎骨-植入物界面,提示螺钉松动问题的风险降低。