Image artifact in dental cone-beam CT.

作者信息

Katsumata Akitoshi, Hirukawa Akiko, Noujeim Marcel, Okumura Shinji, Naitoh Munetaka, Fujishita Masami, Ariji Eiichiro, Langlais Robert P

机构信息

Department of Oral Radiology, Asahi University School of Dentistry, Gifu, Japan.

出版信息

Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2006 May;101(5):652-7. doi: 10.1016/j.tripleo.2005.07.027. Epub 2005 Nov 11.

Abstract

PURPOSE

The purpose of this study was to investigate the appearance and possible cause of an artifact seen in limited-volume cone-beam CT imaging.

METHODS

A water-filled plastic cylinder was used as a phantom of the head. A test object was constructed as a bone-equivalent phantom to be imaged. The test object was variously positioned at the center of the phantom and near its margins. CT images of the test object were acquired using a 3DX Accuitomo system.

RESULTS

In slice images with the test object positioned near the margin of the phantom, arch-shaped defects or deformities were observed on the side of the object. There was a negative correlation between the artifact and the CT value of the object. The artifact was larger in images scanned with a higher voltage.

CONCLUSION

The probability that this artifact is caused by halation from the image intensifier (II) system is suggested.

摘要

目的

本研究旨在调查在有限体积锥形束CT成像中出现的一种伪影的表现及可能原因。

方法

使用一个装满水的塑料圆柱体作为头部模型。构建一个骨等效模型作为待成像的测试物体。将测试物体分别置于模型中心及其边缘附近。使用3DX Accuitomo系统获取测试物体的CT图像。

结果

在测试物体位于模型边缘附近的切片图像中,在物体一侧观察到拱形缺陷或变形。伪影与物体的CT值之间呈负相关。在较高电压扫描的图像中伪影更大。

结论

提示该伪影由图像增强器(II)系统的光晕引起的可能性。