An algorithm for real-time dosimetry in intensity-modulated radiation therapy using the radioluminescence signal from Al2O3:C.

作者信息

Andersen C E, Marckmann C J, Aznar M C, Bøtter-Jensen L, Kjaer-Kristoffersen F, Medin J

机构信息

Risø National Laboratory, Radiation Research Department, DK-4000 Roskilde, Denmark.

出版信息

Radiat Prot Dosimetry. 2006;120(1-4):7-13. doi: 10.1093/rpd/nci600. Epub 2006 Apr 27.

Abstract

Although the radioluminescence (RL) signal from optical fibre Al(2)O(3):C dosemeters used in medical applications is essentially proportional to dose rate, the crystals used so far are imperfect in the sense that their RL sensitivity changes with accumulated dose. A computational algorithm has been developed that corrects for these sensitivity changes. We further report on a new system that effectively separates the RL signal generated in the crystal from fluorescence and Cerenkov emission generated in the optical fibre cable using a gating technique in connection with pulsed linear accelerator radiation beams. The dosimetry system has been used for dose measurements in a phantom during an intensity-modulated radiation therapy (IMRT) treatment with 6 MV photons. The RL measurement results are in excellent agreement (i.e. within 1%) with both the OSL results and the dose delivered according to the treatment planning system. RL signals from Al(2)O(3):C can be used for real-time dose rate measurements with a time resolution of approximately 0.1 s and a spatial resolution only limited by the size of the detector (<0.5 mm).

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验