Yew A, Jin Z M, Donn A, Morlock M M, Isaac G
Division of Bioengineering, National University of Singapore, Singapore.
Proc Inst Mech Eng H. 2006 Feb;220(2):311-9. doi: 10.1243/095441105X69105.
The deformation of metallic acetabular cups employed for metal-on-metal hip resurfacing procedures was considered theoretically using the finite element method in the present study, following on the experimental investigation reported in Part 1. Three representative cups, characterized by the cup wall thickness as thin, intermediate, and thick, were considered. For the intermediate cup, the effects of both the size and the diametral interference on the cup deformation were investigated. Both two-dimensional axisymmetric and three-dimensional finite element models were developed to examine the important parameters during and after the press-fit procedure, and in particular the deformation of the metallic cup. The theoretical prediction of the cup deformation was in reasonable agreement with the corresponding experimental measurement reported in Part 1. The most significant factor influencing the cup deformation was the cup wall thickness. Both the size and the diametral interference were also shown to influence the cup deformation. It is important to ensure that the cup deformation does not significantly affect the clearance designed and optimized for tribological performances of metal-on-metal hip resurfacing prostheses. Furthermore the contact parameters at the cup and bone interface associated with the press fit were also discussed.
在本研究中,继第1部分报道的实验研究之后,采用有限元方法从理论上考虑了用于金属对金属髋关节表面置换手术的金属髋臼杯的变形情况。研究了三种具有代表性的髋臼杯,分别以杯壁厚度薄、中等和厚为特征。对于中等厚度的髋臼杯,研究了尺寸和直径干涉对杯体变形的影响。开发了二维轴对称和三维有限元模型,以研究压配过程中和压配后的重要参数,特别是金属杯的变形。杯体变形的理论预测与第1部分报道的相应实验测量结果合理吻合。影响杯体变形的最主要因素是杯壁厚度。尺寸和直径干涉也被证明会影响杯体变形。确保杯体变形不会显著影响为金属对金属髋关节表面置换假体的摩擦学性能而设计和优化的间隙非常重要。此外,还讨论了与压配相关的杯体与骨界面处的接触参数。