Ushe Mwiza, Mink Jonathan W, Tabbal Samer D, Hong Minna, Schneider Gibson Patricia, Rich Keith M, Lyons Kelly E, Pahwa Rajesh, Perlmutter Joel S
Department of Neurology, Washington University, St. Louis, Missouri, USA.
Mov Disord. 2006 Aug;21(8):1290-2. doi: 10.1002/mds.20926.
Deep brain stimulation (DBS) of the ventral intermediate nucleus (VIM) reduces tremor in people with essential tremor (ET), yet the dependence of tremor suppression on stimulation frequency remains unclear. To address this issue, we tested tremor suppression for three 15-second measurements during a variety of stimulation frequencies in 11 ET patients treated with VIM DBS. Stimulation frequencies at or above 100 Hz produced maximal benefit; higher frequencies provided no additional benefit. If this short-term measure predicts long-term response in routine activities at home, then this stimulation frequency setting will prolong battery half-life compared to higher frequency settings. These findings suggest that ET patients treated with VIM DBS may receive adequate benefit from stimulation frequencies about 100 Hz and this setting compared to commonly used higher settings will prolong battery life of surgically implanted pulse generators.
腹中间核(VIM)的深部脑刺激(DBS)可减轻特发性震颤(ET)患者的震颤,但震颤抑制对刺激频率的依赖性仍不清楚。为解决这一问题,我们对11例接受VIM DBS治疗的ET患者在多种刺激频率下进行了三次15秒测量,以测试震颤抑制情况。100Hz及以上的刺激频率产生了最大益处;更高的频率没有额外益处。如果这种短期测量能预测在家中日常活动中的长期反应,那么与更高频率设置相比,这种刺激频率设置将延长电池半衰期。这些发现表明,接受VIM DBS治疗的ET患者可能从约100Hz的刺激频率中获得足够益处,与常用的更高设置相比,这种设置将延长手术植入脉冲发生器的电池寿命。