[波兰使用胶片增感箔系统降低乳腺摄影辐射剂量]

[Reduction of radiation dosage in mammography by using film intensifying foil systems in Poland].

作者信息

Iwaszkiewicz K, Bończyk J

机构信息

Zakładu Rentgenodiagnostyki Instytutu Radiologii AM, Laboratorium Radiologii WZF Foton, Warszawie.

出版信息

Pol Tyg Lek. 1991;46(17-18):346-8.

PMID:1669071
Abstract

In order to assess possibility of radiation dose reduction during mammography at direct film radiation, a comparative studies have been carried out. The studies involved Roentgen XT-2 film without intensifying foil compared with Mammoray MR-3 (Agfa Gevaert), Ortho M-MO-1 (Kodak) in the original cassettes, and Fluorfilm TK-1 (WZF Foton) with foils Min.-R. Film-intensifying foil systems are four and six times more sensitive than Rentgen XT-2 film. Sensitometric and diagnostic properties of Fluorofilm TK-1 are comparable with Mammoray MR-3 film. Therefore, films Rentgen XT-2 which are presently used in mammography should be replaced with Fluorofilm TK-1 with the intensifying foils.

摘要

为了评估直接胶片放射乳腺摄影时降低辐射剂量的可能性,已开展了一项对比研究。该研究涉及未使用增感箔的Roentgen XT - 2胶片,并与原装暗盒中的Mammoray MR - 3(爱克发吉华)、Ortho M - MO - 1(柯达)进行比较,以及使用Min.-R箔的Fluorfilm TK - 1(WZF Foton)。胶片 - 增感箔系统比Roentgen XT - 2胶片敏感4至6倍。Fluorofilm TK - 1的感光测定和诊断性能与Mammoray MR - 3胶片相当。因此,目前乳腺摄影中使用的Roentgen XT - 2胶片应被带有增感箔的Fluorofilm TK - 1所取代。