Aslund Magnus, Cederström Björn, Lundqvist Mats, Danielsson Mats
Department of Physics, Royal Institute of Technology, AlbaNova, 106 91, Stockholm, Sweden.
Med Phys. 2006 Apr;33(4):933-40. doi: 10.1118/1.2179122.
The scatter to primary ratio (SPR) was measured on a scanning multislit full-field digital mammography system for different thickness of breast equivalent material and different tube voltages. Scatter within the detector was measured separately and was found to be the major source of scatter in the assembly. Measured total SPRs below 6% are reported for breast range 3-7 cm. The performance of the multislit assembly is compared to other imaging geometries with different scatter rejection schemes by using the scatter detective quantum efficiency.
在扫描多狭缝全场数字化乳腺摄影系统上,针对不同厚度的等效乳腺材料和不同管电压测量了散射与原发射线比值(SPR)。分别测量了探测器内的散射,发现其是该组件中散射的主要来源。报告了乳腺厚度在3 - 7厘米范围内测得的总SPR低于6%的情况。通过使用散射探测量子效率,将多狭缝组件的性能与具有不同散射抑制方案的其他成像几何结构进行了比较。