Padley D, Strauss R G, Wieland M, Randels M J
Elmer L. DeGowin Memorial Blood Center, University of Iowa Hospitals and Clinics, University of Iowa College of Medicine, Iowa City 52242.
J Clin Apher. 1991;6(2):77-80. doi: 10.1002/jca.2920060203.
Hematopoietic stem cells, collected by leukapheresis from peripheral blood, can be used as an alternative to autologous bone marrow transplantation following high-dose chemotherapy as treatment of several malignancies. We compared the ability of the Cobe Spectra and the Fenwal CS3000 to collect peripheral blood mononuclear cells (MNC) for autologous peripheral blood stem cell transplantation. Ten patients experienced repeated leukapheresis (10 L blood processed per procedure) using both instruments. Procedures were alternated between the two until a total of 7 x 10(8) MNC/kg was collected. Data from 61 Spectra and 50 CS3000 collections were analyzed. The yield (mean per procedure) of nucleated cells (NC) and MNC was higher (P less than .005) with the Spectra (0.77 x 10(10) NC and 0.54 x 10(10) MNC) than with the CS3000 (0.59 x 10(10) NC and 0.40 x 10(10) MNC). However, colony forming units (CFU-GM) were not different (P greater than .05) for Spectra (0.92 x 10(4)) and Fenwal (0.65 x 10(4) collections. Platelet contamination was lower (P less than .001) with the Spectra (2.2 x 10(11)) than the CS3000 (5.0 x 10(11)). This correlated with a higher patient blood platelet count immediately following Spectra collections (117 x 10(9)/L) versus the CS3000 (86 x 10(9)/L). Using the methods described, the Spectra product contained greater yields of NC and MNC with less platelet contamination than did the CS3000.
通过白细胞单采术从外周血中采集的造血干细胞,可作为高剂量化疗后自体骨髓移植的替代方法,用于治疗多种恶性肿瘤。我们比较了Cobe Spectra和Fenwal CS3000采集外周血单个核细胞(MNC)用于自体外周血干细胞移植的能力。10例患者使用这两种仪器进行了重复白细胞单采术(每次处理10L血液)。两种仪器的操作交替进行,直至共采集到7×10⁸MNC/kg。分析了61次Spectra采集和50次CS3000采集的数据。Spectra采集的有核细胞(NC)和MNC产量(每次操作的平均值)高于CS3000(P<0.005),分别为0.77×10¹⁰NC和0.54×10¹⁰MNC,而CS3000分别为0.59×10¹⁰NC和0.40×10¹⁰MNC。然而,Spectra(0.92×10⁴)和Fenwal(0.65×10⁴)采集的集落形成单位(CFU-GM)没有差异(P>0.05)。Spectra采集后的血小板污染低于CS3000(P<0.001),分别为2.2×10¹¹和5.0×10¹¹。这与Spectra采集后患者的血小板计数(117×10⁹/L)高于CS3000采集后(86×10⁹/L)相关。使用所述方法,Spectra采集产物的NC和MNC产量更高,血小板污染更少,优于CS3000。