Field comparison of commercially available short-term radon detectors.

作者信息

Sun Kainan, Majdan Marek, Field Daniel W, Field R William

机构信息

College of Public Health, 104 IREH, University of Iowa, Iowa City, IA 52242, USA.

出版信息

Health Phys. 2006 Sep;91(3):221-6. doi: 10.1097/01.HP.0000206040.55298.ce.

Abstract

We performed a comparison of commercially available short-term radon detectors in order to determine the accuracy and precision of the detectors under actual field conditions. We exposed fifteen radon detectors, under field conditions, from each of six companies to a reference radon concentration. Five of the six companies tested did not pass the U.S. Environmental Protection Agency's previously established accuracy guideline (all individual relative errors </= 25%) and four of the six companies failed the EPA's precision guideline [coefficient of variation </= 10% at 150 Bq m (4 pCi L)]. The findings suggest that the performances of commercially available radon detectors exposed under actual field conditions may not be as accurate or precise as those detectors available prior to the close of the EPA's National Radon Proficiency Program in 1998. It is unknown if this one-time "snap shot" represents the overall reliability of the accuracy and precision of commercially available radon detectors. Nonetheless, the findings suggest that additional double-blind testing of commercially available radon detectors under actual field conditions is warranted.

摘要

我们对市售的短期氡探测器进行了比较,以确定这些探测器在实际现场条件下的准确性和精密度。我们在现场条件下,将来自六家公司的15个氡探测器暴露于参考氡浓度环境中。测试的六家公司中有五家未通过美国环境保护局先前制定的准确性指南(所有个体相对误差≤25%),六家公司中有四家未通过美国环境保护局的精密度指南[在150 Bq/m³(4 pCi/L)时变异系数≤10%]。研究结果表明,在实际现场条件下暴露的市售氡探测器的性能可能不如1998年美国环境保护局国家氡熟练度计划结束前的那些探测器准确或精密。尚不清楚这一“快照”是否代表了市售氡探测器准确性和精密度的整体可靠性。尽管如此,研究结果表明有必要在实际现场条件下对市售氡探测器进行额外的双盲测试。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验