Chen Albert P, Cunningham Charles H, Kurhanewicz John, Xu Duan, Hurd Ralph E, Pauly John M, Carvajal Lucas, Karpodinis Kostas, Vigneron Daniel B
Department of Radiology, University of California at San Francisco, San Francisco, CA 94143-2512, USA.
Magn Reson Imaging. 2006 Sep;24(7):825-32. doi: 10.1016/j.mri.2006.03.002. Epub 2006 May 2.
A 3 T MLEV-point-resolved spectroscopy (PRESS) sequence employing optimized spectral-spatial and very selective outer-voxel suppression pulses was tested in 25 prostate cancer patients. At an echo time of 85 ms, the MLEV-PRESS sequence resulted in maximally upright inner resonances and minimal outer resonances of the citrate doublet of doublets. Magnetic resonance spectroscopic imaging (MRSI) exams performed at both 3 and 1.5 T for 10 patients demonstrated a 2.08+/-0.36-fold increase in signal-to-noise ratio (SNR) at 3 T as compared with 1.5 T for the center citrate resonances. This permitted the acquisition of MRSI data with a nominal spatial resolution of 0.16 cm3 at 3 T with similar SNR as the 0.34-cm3 data acquired at 1.5 T. Due to the twofold increase in spectral resolution at 3 T and the improved magnetic field homogeneity provided by susceptibility-matched endorectal coils, the choline resonance was better resolved from polyamine and creatine resonances as compared with 1.5 T spectra. In prostate cancer patients, the elevation of choline and the reduction of polyamines were more clearly observed at 3 T, as compared with 1.5 T MRSI. The increased SNR and corresponding spatial resolution obtainable at 3 T reduced partial volume effects and allowed improved detection of the presence and extent of abnormal metabolite levels in prostate cancer patients, as compared with 1.5 T MRSI.
在25例前列腺癌患者中测试了一种采用优化的频谱空间和非常选择性的体外体素抑制脉冲的3T MLEV点分辨光谱(PRESS)序列。在回波时间为85毫秒时,MLEV-PRESS序列使双重峰柠檬酸盐的内部共振最大化,外部共振最小化。对10例患者在3T和1.5T下进行的磁共振波谱成像(MRSI)检查表明,与1.5T相比,3T时中心柠檬酸盐共振的信噪比(SNR)提高了2.08±0.36倍。这使得在3T时能够采集标称空间分辨率为0.16cm³的MRSI数据,其SNR与在1.5T时采集的0.34cm³数据相似。由于3T时光谱分辨率提高了两倍,以及由易感性匹配的直肠内线圈提供的磁场均匀性改善,与1.5T光谱相比,胆碱共振与多胺和肌酸共振的分辨更好。在前列腺癌患者中,与1.5T MRSI相比,在3T时胆碱升高和多胺降低的情况更明显。与1.5T MRSI相比,3T时可获得的SNR增加和相应的空间分辨率降低了部分容积效应,并改善了对前列腺癌患者异常代谢物水平的存在和范围的检测。