Suppr超能文献

为了获得可靠的自引导扩散张量磁共振成像,究竟需要收集多少数据?

Just how much data need to be collected for reliable bootstrap DT-MRI?

作者信息

O'Gorman Ruth L, Jones Derek K

机构信息

Department of Neuroimaging, King's College Hospital, London, UK.

出版信息

Magn Reson Med. 2006 Oct;56(4):884-90. doi: 10.1002/mrm.21014.

Abstract

Diffusion tensor MRI (DT-MRI) can provide estimates of fiber orientation derived from the orientational dependence of the diffusivity of water molecules, enabling the reconstruction of white matter fiber pathways using tractography methods. However, noise arising from various sources can introduce uncertainty into the estimates of the elements of the diffusion tensor, resulting in errors in fiber orientation estimates such that tractography reconstructions of fiber pathways potentially can be imprecise and inaccurate. Recently, attempts have been made to characterize the uncertainty in DT-MRI-derived parameters using the bootstrap method; however, several questions remain open regarding the number of repeat measurements and bootstraps required to accurately and precisely reconstruct the probability distributions of the DT-MRI parameters. This study investigates the accuracy and precision of the bootstrap method for characterizing distributions of DT-MRI parameters. A number of experimental bootstrap designs and sampling schemes containing different numbers of isotropically distributed gradient vectors are considered, using an idealized system where the true variability in each parameter is known. This study demonstrates that for most DT-MRI experiments, robust results will be obtained if the minimum number of bootstraps is approximately 500, and that at least five repeat samples of each diffusion-weighted intensity should be used for bootstrapping.

摘要

扩散张量磁共振成像(DT-MRI)可以根据水分子扩散率的方向依赖性提供纤维方向的估计值,从而能够使用纤维束成像方法重建白质纤维通路。然而,来自各种来源的噪声会给扩散张量元素的估计值带来不确定性,导致纤维方向估计出现误差,使得纤维通路的纤维束成像重建可能不准确且不精确。最近,人们尝试使用自助法来表征DT-MRI衍生参数中的不确定性;然而,关于准确且精确地重建DT-MRI参数的概率分布所需的重复测量次数和自助抽样次数,仍有几个问题尚未解决。本研究调查了用于表征DT-MRI参数分布的自助法的准确性和精确性。使用一个理想化系统,其中每个参数的真实变异性是已知的,考虑了许多包含不同数量各向同性分布梯度向量的实验自助设计和抽样方案。本研究表明,对于大多数DT-MRI实验,如果自助抽样的最小次数约为500,将获得稳健的结果,并且每个扩散加权强度至少应使用五个重复样本进行自助抽样。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验