Comparison of corneal thickness measured with optical coherence tomography, ultrasonic pachymetry, and a scanning slit method.

作者信息

Thomas Joseph, Wang Jianhua, Rollins Andrew M, Sturm Julie

机构信息

Department of Ophthalmology, Case Western Reserve University/University Hospitals of Cleveland, Cleveland, Ohio, USA.

出版信息

J Refract Surg. 2006 Sep;22(7):671-8. doi: 10.3928/1081-597X-20060901-08.

Abstract

PURPOSE

To investigate the relationship between 1310-nm optical coherence tomography (OCT), an ultrasound pachymeter, and the Orbscan II (Bausch & Lomb, Rochester, NY) in corneal thickness measurements before and after LASIK.

METHODS

Twenty-eight eyes of 14 patients with a mean age of 39.9 +/- 8.6 years underwent LASIK for correction of myopia with or without astigmatism. Central corneal thickness was measured with real-time 1310-nm OCT and Orbscan preoperatively and 1 day, 1 week, and 1 month postoperatively, and with an ultrasound pachymeter preoperatively and 1 month postoperatively.

RESULTS

Measurements of corneal thickness by all instruments correlated well with one another pre- and postoperatively (r value range: 0.87 to 0.94, P<.05); however, there were significant differences among these methods (ANOVA: P<.005). Baseline readings for ultrasound and Orbscan were similar, as were readings 1 month postoperatively for OCT and Orbscan (P>.05). Thickness measured by OCT was thinner than for other instruments preoperatively (P<.005) and thicker than Orbscan postoperatively (P<.005). Significant corneal thinning (P<.001) was found with all study devices 1 day after laser ablation. Significant corneal thinning (P<.005) was found with OCT and Orbscan 1 week compared to 1 day postoperatively, and thickening (post hoc: P<.001) was evident at 1 month compared to 1 week postoperatively.

CONCLUSIONS

Although the instruments correlated well in measuring corneal thickness after LASIK, there were significant differences. All instruments indicated corneal swelling and recovery during the first postoperative week and thickening at 1 month.

摘要

目的

研究1310纳米光学相干断层扫描(OCT)、超声角膜测厚仪与Orbscan II(博士伦公司,纽约州罗切斯特)在准分子原位角膜磨镶术(LASIK)前后角膜厚度测量中的关系。

方法

14例平均年龄为39.9±8.6岁的患者的28只眼睛接受了LASIK手术,用于矫正近视合并或不合并散光。术前以及术后1天、1周和1个月,使用实时1310纳米OCT和Orbscan测量中央角膜厚度,术前和术后1个月使用超声角膜测厚仪测量。

结果

所有仪器测量的角膜厚度在术前和术后彼此相关性良好(r值范围:0.87至0.94,P<0.05);然而,这些方法之间存在显著差异(方差分析:P<0.005)。超声和Orbscan的基线读数相似,OCT和Orbscan术后1个月的读数也相似(P>0.05)。术前OCT测量的厚度比其他仪器薄(P<0.005),术后比Orbscan厚(P<0.005)。激光消融术后1天,所有研究设备均发现角膜显著变薄(P<0.001)。与术后1天相比,OCT和Orbscan在术后1周发现角膜显著变薄(P<0.005),与术后1周相比,术后1个月增厚明显(事后检验:P<0.001)。

结论

尽管这些仪器在LASIK术后测量角膜厚度时相关性良好,但仍存在显著差异。所有仪器均显示术后第一周角膜肿胀和恢复,术后1个月增厚。