Ho Thomas, Cheng Arthur C K, Rao Srinivas K, Lau Silvania, Leung Chris K S, Lam Dennis S C
Department of Ophthalmology & Visual Sciences, The Chinese University of Hong Kong, University Eye Center, Hong Kong Eye Hospital, Hong Kong, China.
J Cataract Refract Surg. 2007 Jul;33(7):1177-82. doi: 10.1016/j.jcrs.2007.03.028.
To compare corneal pachymetry assessment using 4 measurement methods in eyes after laser in situ keratomileusis (LASIK) for myopia.
Department of Ophthalmology & Visual Sciences, The Chinese University of Hong Kong, Hong Kong Eye Hospital, Hong Kong SAR.
Fifty-two consecutive patients (103 eyes) who had LASIK for the correction of myopia had Orbscan II (Bausch & Lomb), Visante (Carl Zeiss Meditec), Pentacam (Oculus, Inc.), and ultrasound (US) pachymetry (Sonomed, 200P) 6 months after surgery. Data were analyzed using the paired sample t test, Bland-Altman plots, and linear regression.
The mean postoperative pachymetry measured by US, Orbscan (0.89 acoustic factor), Pentacam, and Visante pachymetry were 438.2 microm+/-41.18 (SD), 435.17+/-49.63 microm, 430.66+/-40.23 microm, and 426.56+/-41.6 microm, respectively. Compared with the US measurement, Pentacam and Visante measurements significantly underestimated corneal thickness by a mean of 7.54+/-15.06 microm (P<.01) and 11.64+/-12.87 microm (P<.01), respectively. There was no statistically significant difference between US and Orbscan measurements.
Pentacam and Visante measurements of corneal thickness 6 months after LASIK were significantly less than those obtained using Orbscan and US pachymetry, although all 4 measurement methods showed a high correlation with each other.
比较准分子原位角膜磨镶术(LASIK)治疗近视后使用4种测量方法评估角膜厚度的情况。
香港特别行政区香港中文大学眼科及视觉科学系、香港眼科医院。
52例连续行LASIK矫正近视的患者(103只眼)在术后6个月接受Orbscan II(博士伦公司)、Visante(卡尔蔡司医疗技术公司)、Pentacam(奥culus公司)及超声(US)角膜测厚(Sonomed,200P)检查。采用配对样本t检验、Bland-Altman图及线性回归分析数据。
US、Orbscan(0.89声学因子)、Pentacam及Visante测量的术后平均角膜厚度分别为438.2微米±41.18(标准差)、435.17±49.63微米、430.66±40.23微米及426.56±41.6微米。与US测量结果相比,Pentacam和Visante测量结果分别显著低估角膜厚度,平均低估7.54±15.06微米(P<0.01)和11.64±12.87微米(P<0.01)。US和Orbscan测量结果之间无统计学显著差异。
LASIK术后6个月,Pentacam和Visante测量的角膜厚度显著低于Orbscan和US角膜测厚结果,尽管所有4种测量方法之间均显示出高度相关性。