Muacevic Alexander, Staehler Michael, Drexler Christian, Wowra Berndt, Reiser Maximilian, Tonn Joerg-Christian
European CyberKnife Center Munich, Germany.
J Neurosurg Spine. 2006 Oct;5(4):303-12. doi: 10.3171/spi.2006.5.4.303.
The authors describe the technical application of the Xsight Spine Tracking System, data pertaining to accuracy obtained during phantom testing, and the initial clinical feasibility of using this fiducial-free alignment system with the CyberKnife in spinal radiosurgery.
The Xsight integrates with the CyberKnife radiosurgery system to eliminate the need for implantation of radiographic markers or fiducials prior to spinal radiosurgery. It locates and tracks spinal lesions relative to spinal osseous landmarks. The authors performed 10 end-to-end tests of accuracy using an anthropomorphic head and cervical spine phantom. Xsight was also used in the treatment of 50 spinal lesions in 42 patients. Dose planning was based on 1.5-mm-thick computed tomography slices in which an inverse treatment planning technique was used. All lesions could be treated using the fiducial-free tracking procedure. Phantom tests produced an overall mean targeting error of 0.52 +/- 0.22 mm. The setup time for patient alignment averaged 6 minutes (range 2-45 minutes). The treatment doses varied from 12 to 25 Gy to the median prescription isodose of 65% (40 to 70%). The tumor volume ranged between 1.3 and 152.8 cm3. The mean spinal cord volume receiving greater than 8 Gy was 0.69 +/- 0.35 cm3. No short-term adverse events were noted during the 1- to 7-month follow-up period. Axial and radicular pain was relieved in 14 of 15 patients treated for pain.
Fiducial-free tracking is a feasible, accurate, and reliable tool for radiosurgery of the entire spine. By eliminating the need for fiducial implantation, the Xsight system offers patients noninvasive radiosurgical intervention for intra- and paraspinal tumors.
作者描述了Xsight脊柱追踪系统的技术应用、在体模测试中获得的与准确性相关的数据,以及在脊柱放射外科手术中使用这种无标记对准系统与射波刀联合的初步临床可行性。
Xsight与射波刀放射外科系统集成,消除了脊柱放射外科手术前植入放射学标记物或基准物的需求。它相对于脊柱骨性标志定位并追踪脊柱病变。作者使用拟人化头部和颈椎体模进行了10次端到端的准确性测试。Xsight还用于治疗42例患者的50处脊柱病变。剂量规划基于1.5毫米厚的计算机断层扫描切片,采用逆向治疗规划技术。所有病变均可使用无标记追踪程序进行治疗。体模测试产生的总体平均靶向误差为0.52±0.22毫米。患者对准的设置时间平均为6分钟(范围为2至45分钟)。治疗剂量从12至25 Gy不等,中位处方等剂量线为65%(40至70%)。肿瘤体积在1.3至152.8立方厘米之间。接受大于8 Gy照射的平均脊髓体积为0.69±0.35立方厘米。在1至7个月的随访期内未观察到短期不良事件。15例因疼痛接受治疗的患者中有14例的轴向和神经根性疼痛得到缓解。
无标记追踪是全脊柱放射外科手术可行、准确且可靠的工具。通过消除植入基准物的需求,Xsight系统为患者提供了针对椎内和椎旁肿瘤的非侵入性放射外科干预。