Suppr超能文献

金属对金属全髋关节置换术和髋关节表面置换术中铬、钴和钼的暴露情况。

Exposure to chromium, cobalt and molybdenum from metal-on-metal total hip replacement and hip resurfacing arthroplasty.

作者信息

Witzleb Wolf-Christoph, Ziegler Joerg, Krummenauer Frank, Neumeister Volker, Guenther Klaus-Peter

机构信息

Department of Orthopaedic Surgery, Universty Hospital Carl Gustav Carus, Dresden University of Technology, Dresden, Germany.

出版信息

Acta Orthop. 2006 Oct;77(5):697-705. doi: 10.1080/17453670610012863.

Abstract

BACKGROUND

All metal implants--and metal-on-metal bearings in particular--corrode and cause a release of metal ions. Because cobalt and chromium have been shown to be carcinogenic and mutagenic in human and animal models, systemic toxicity and cancer risk are considered to be possible disadvantages of the metal-on-metal articulation. This study was designed to investigate the serum concentration profiles of chromium, cobalt and molybdenum after implantation of a Birmingham hip resurfacing arthroplasty (BHR) and a cementless total hip replacement with a 28-mm Metasul articulation (MTHR), over the first 2 years after implantation.

METHODS

We analyzed profiles of metal ion serum levels in 111 patients implanted with a BHR, in 74 patients implanted with an MTHR, and in 130 implant-free probands control subjects using atomic absorption spectrophotometry.

RESULTS

Chromium and cobalt concentrations (in microg/L) of all BHR and MTHR patients differed significantly from those of control subjects (chromium: < 0.25; cobalt: 0.25). The median chromium and cobalt concentrations in BHR patients had increased to 5.1 and 4.3 microg/L 2 years after surgery. Concentrations in BHR patient exceeded those in the unilateral MTHR patients. Molybdenum serum concentrations hardly changed over time in either group and were not significantly different from the concentrations seen in the control subjects.

INTERPRETATION

During the first 2 years after surgery, the Birmingham hip resurfacing arthroplasty leads to a significantly greater increase in serum chromium and cobalt levels than the 28-mm metal-on-metal MTHR. Observation of patients over a longer period will be necessary in order to evaluate any chronic adverse effects to the system due to elevated chromium and cobalt serum concentrations.

摘要

背景

所有金属植入物,尤其是金属对金属的关节轴承,都会发生腐蚀并导致金属离子释放。由于钴和铬在人类和动物模型中已被证明具有致癌性和致突变性,全身毒性和癌症风险被认为是金属对金属关节置换的潜在缺点。本研究旨在调查植入伯明翰髋关节表面置换术(BHR)和采用28毫米Metasul关节的非骨水泥全髋关节置换术(MTHR)后,在植入后的头两年内铬、钴和钼的血清浓度变化情况。

方法

我们使用原子吸收分光光度法分析了111例接受BHR植入的患者、74例接受MTHR植入的患者以及130例无植入物的对照受试者的金属离子血清水平变化情况。

结果

所有BHR和MTHR患者的铬和钴浓度(微克/升)与对照受试者有显著差异(铬:<0.25;钴:0.25)。BHR患者术后2年铬和钴的中位浓度分别增至5.1和4.3微克/升。BHR患者的浓度超过单侧MTHR患者。两组患者钼的血清浓度随时间变化不大,与对照受试者的浓度无显著差异。

解读

在术后的头两年内,伯明翰髋关节表面置换术导致血清铬和钴水平的升高幅度明显大于28毫米金属对金属的MTHR。为了评估因血清铬和钴浓度升高对系统产生的任何慢性不良影响,有必要对患者进行更长时间的观察。