Luetzner Joerg, Krummenauer Frank, Lengel Attila Michael, Ziegler Joerg, Witzleb Wolf-Christoph
Department of Orthopaedic Surgery, University Hospital Carl Gustav Carus, Dresden, Germany.
Clin Orthop Relat Res. 2007 Aug;461:136-42. doi: 10.1097/BLO.0b013e31806450ef.
All metal implants release metal ions because of corrosion. This has been studied and debated, especially in metal-on-metal total hip arthroplasties. Total knee arthroplasty implants have large metal surface areas and therefore substantial potential for corrosion. We determined changes in serum levels of metal ions in 41 patients after cemented unconstrained total knee arthroplasty without patellar resurfacing, 18 with unilateral total knee arthroplasty (median, 66 months after surgery) and 23 patients with bilateral total knee arthroplasties (75 and 50 months after first and second surgeries, respectively). Serum concentrations of chromium, cobalt, and molybdenum were analyzed and related to the number of total knee arthroplasties and compared with those of 130 control patients without implants. The median chromium, cobalt, and molybdenum concentrations were 0.92, 3.28, and 2.55 microg/L, respectively, in the unilateral total knee arthroplasty sample and 0.98, 4.28, and 2.40 microg/L, respectively, in the bilateral total knee arthroplasty sample. We observed no difference between the serum levels in patients with unilateral and bilateral arthroplasties, but the serum levels of chromium and cobalt of both study groups were greater than those of the control group (less than 0.25 microg/L). The patients who had total knee arthroplasty had molybdenum profiles that were similar to those of the control group (median, 2.11 microg/L).
所有金属植入物都会因腐蚀而释放金属离子。这一现象已得到研究和讨论,尤其是在金属对金属全髋关节置换术中。全膝关节置换植入物具有较大的金属表面积,因此存在大量的腐蚀可能性。我们测定了41例行骨水泥固定非限制性全膝关节置换术且未进行髌骨表面置换患者的血清金属离子水平变化,其中18例为单侧全膝关节置换术患者(术后中位时间为66个月),23例为双侧全膝关节置换术患者(分别在首次和第二次手术后75个月和50个月)。分析了血清中铬、钴和钼的浓度,并将其与全膝关节置换术的数量相关联,同时与130例未植入假体的对照患者进行比较。单侧全膝关节置换术样本中铬、钴和钼的中位浓度分别为0.92、3.28和2.55微克/升,双侧全膝关节置换术样本中分别为0.98、4.28和2.40微克/升。我们观察到单侧和双侧置换术患者的血清水平无差异,但两个研究组的血清铬和钴水平均高于对照组(低于0.25微克/升)。接受全膝关节置换术的患者钼水平与对照组相似(中位值为2.11微克/升)。