Inter-instrument agreement and influence of central corneal thickness on measurements with Goldmann, pneumotonometer and noncontact tonometer in glaucomatous eyes.

作者信息

Gupta Viney, Sony Parul, Agarwal Harish C, Sihota Ramanjit, Sharma Ajay

机构信息

Dr. Rajendra Prasad Centre for Ophthalmic Sciences, All India Institute of Medical Sciences, New Delhi - 110 029, India.

出版信息

Indian J Ophthalmol. 2006 Dec;54(4):261-5. doi: 10.4103/0301-4738.27952.

Abstract

PURPOSE

This study was conducted to compare the intraocular pressure (IOP) measurements by the Goldman applanation tonometer (GAT), non contact tonometer (NCT) and the ocular blood flow (OBF) pneumotonometer in different IOP ranges in glaucomatous eyes. The effect of central corneal thickness (CCT) on IOP measurement in chronic glaucomatous eyes using the three different tonometers was also evaluated.

MATERIALS AND METHODS

IOP measurements of 130 eyes of primary glaucoma patients were performed using GAT by an ophthalmologist while NCT and OBF-pneumotonometer measurements were performed by an experienced optometrist. The IOP values were compared amongst the three instruments in the three different IOP ranges (0-18 mmHg, > 18 to 25 mmHg, > 25 mmHg). CCT was also measured in all patients.

RESULTS

The mean of paired difference between GAT and NCT was 0.9 +/- 3.1 mmHg while that between GAT and OBF-pneumotonometer was 0.3 +/- 3.4 mmHg. The OBF-pneumotonometer and NCT were more affected by corneal thickness (0.41 mmHg and 0.4 mmHg / 10 micro corneal thickness respectively) while GAT was the least affected by corneal thickness (0.3 mmHg / 10 micro corneal thickness) though the difference was not statistically significant ( P =0.42).

CONCLUSION

With appropriate correction for corneal thickness the NCT and OBF-pneumotonometer can be used as reliably as GAT in following up glaucomatous patients.