Tonnu P-A, Ho T, Newson T, El Sheikh A, Sharma K, White E, Bunce C, Garway-Heath D
The Children's University Hospital, National Children's Eye Center, and University College Lublin, Republic of Ireland.
Br J Ophthalmol. 2005 Jul;89(7):851-4. doi: 10.1136/bjo.2004.056622.
To evaluate the influence of central corneal thickness (CCT) on intraocular pressure (IOP) measurements made with the Goldmann applanation tonometer (GAT), Tono-Pen XL, ocular blood flow tonograph (OBF), and Canon TX-10 non-contact tonometer (NCT).
CCT was recorded for either eye (randomly selected) of each of 105 untreated patients with ocular hypertension and glaucoma attending the glaucoma research unit at Moorfields Eye Hospital. For each of the selected eyes, IOP was measured with the GAT (two observers), Tono-Pen, OBF, and NCT in a randomised order. The relation of measured IOP and of inter-tonometer differences with CCT and subject age was explored by linear regression analysis.
A significant association between measured IOP and CCT was found with each instrument. The change in measured IOP for a 10 mum increase in CCT was 0.28, 0.31, 0.38, and 0.46 for the GAT, Tono-Pen, OBF, and NCT, respectively (all p< or = 0.05). There was a significant association between the NCT/GAT differences and CCT, with a tendency of NCT to overestimate GAT in eyes with thicker corneas. There was a significant association between GAT/Tono-Pen and OBF/Tono-Pen differences and age, with a tendency of GAT and OBF to overestimate the Tono-Pen in eyes of older subjects.
IOP measurement by all four methods is affected by CCT. The NCT is affected by CCT significantly more than the GAT. Subject age has a differential effect on the IOP measurements made by the GAT and OBF compared to the Tono-Pen.
评估中央角膜厚度(CCT)对使用戈德曼压平眼压计(GAT)、托诺笔XL眼压计、眼血流眼压描记仪(OBF)和佳能TX - 10非接触眼压计(NCT)测量眼压(IOP)的影响。
记录了在摩尔菲尔德眼科医院青光眼研究单元就诊的105例未经治疗的高眼压症和青光眼患者每只眼睛(随机选择)的CCT。对于每只选定的眼睛,随机顺序使用GAT(两名观察者)、托诺笔、OBF和NCT测量眼压。通过线性回归分析探讨测量的眼压以及眼压计间差异与CCT和受试者年龄的关系。
使用每种仪器均发现测量的眼压与CCT之间存在显著关联。CCT每增加10μm,GAT、托诺笔、OBF和NCT测量的眼压变化分别为0.28、0.31、0.38和0.46(所有p≤0.05)。NCT/GAT差异与CCT之间存在显著关联,在角膜较厚的眼睛中,NCT有高估GAT的趋势。GAT/托诺笔和OBF/托诺笔差异与年龄之间存在显著关联,在老年受试者的眼睛中,GAT和OBF有高估托诺笔测量值的趋势。
所有四种方法测量眼压均受CCT影响。NCT受CCT的影响明显大于GAT。与托诺笔相比,受试者年龄对GAT和OBF测量眼压有不同影响。