Abstract

INTRODUCTION

Goldmann applanation tonometer (GAT) is the gold standard for Intraocular Pressure (IOP) measurement but has disadvantage of being contact device and problems with portability. The aim of the study was to compare the Keeler's Pulsair noncontact tonometer (NCT) with GAT in Indian Population.

MATERIALS AND METHODS

Eighty-one subjects were screened from a Glaucoma clinic of a tertiary care centre in North India. The IOP was measured by Pulsair NCT and GAT after explaining the procedure. Central corneal thickness (CCT) was measured to avoid its bias on IOP readings. The data were analyzed using SPSS software.

RESULTS

The mean age of subjects was 49.9±8.8 (mean±SD) years. The mean IOP as taken by Pulsair NCT was 15.79±4.07mmHg and that for GAT was 17.02±4.23mmHg (p=0.062). The mean CCT was 0.536±0.019mm. A positive Pearson's correlation coefficient of 0.909 (p=0.0001) was found between the two instruments. Bland and Altmann analysis showed a fair agreement between the two tonometers at lower IOP range.

CONCLUSION

Pulsair NCT can be used as a screening tool for community practices but is not reliable in the subjects with higher IOP range.

摘要

引言

Goldmann压平眼压计(GAT)是测量眼压(IOP)的金标准,但它是一种接触式设备,存在便携性问题。本研究的目的是在印度人群中比较Keeler的Pulsair非接触眼压计(NCT)与GAT。

材料与方法

从印度北部一家三级护理中心的青光眼诊所筛选出81名受试者。在解释测量过程后,使用Pulsair NCT和GAT测量眼压。测量中央角膜厚度(CCT)以避免其对眼压读数的偏差。使用SPSS软件对数据进行分析。

结果

受试者的平均年龄为49.9±8.8(均值±标准差)岁。Pulsair NCT测得的平均眼压为15.79±4.07mmHg,GAT测得的平均眼压为17.02±4.23mmHg(p = 0.062)。平均CCT为0.536±0.019mm。两种仪器之间的Pearson相关系数为0.909(p = 0.0001)。Bland和Altmann分析表明,在较低眼压范围内,两种眼压计之间具有较好的一致性。

结论

Pulsair NCT可作为社区实践的筛查工具,但在眼压较高的受试者中不可靠。

相似文献

引用本文的文献

5
Neural Mechanisms of Selective Attention in Children with Amblyopia.弱视儿童选择性注意的神经机制
PLoS One. 2015 Jun 11;10(6):e0125370. doi: 10.1371/journal.pone.0125370. eCollection 2015.

本文引用的文献

5
IOP variations in the sitting and supine positions.在坐姿和仰卧位时的眼内压变化。
J Glaucoma. 2010 Dec;19(9):609-12. doi: 10.1097/IJG.0b013e3181ca7ca5.
7
Tonometry through the ages.眼压测量的历史
Surv Ophthalmol. 2008 Nov-Dec;53(6):568-91. doi: 10.1016/j.survophthal.2008.08.024.

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验