Four butyryl trihexyl phthalate plasticized polyvinyl chloride (BTHC-PVC) containers were compared for storage of leukoreduced platelet concentrates (LR-PC): three ethylene oxide (EtO) sterilized (Gambro, Haemonetics, Fresenius), and one steam-sterilized (Fresenius).
LR-PCs were made from 5 buffy coats and 300 ml Composol additive solution, and leukoreduced by filtration. Four LR-PCs were pooled and subsequently divided over the 4 BTHC-PVC bags to prevent donor-dependent differences, and sampled for in vitro analysis on day 1, 2, 5, 7 and 9.
The pH values on day 9 were (mean +/- SD, n = 10): 7.12 +/- 0.03 (Gambro), 7.12 +/- 0.04 (Haemonetics), and 7.07 +/- 0.09 (Fresenius, EtO-sterilized) (not significantly different), vs. 6.91 +/- 0.12 (Fresenius, steam-sterilized; P < 0.001 vs. all EtO-sterilized bags). LR-PCs stored in the steam-sterilized bag exhibited significantly higher glucose consumption and lactate production (P < 0.001 vs. all EtO-sterilized bags).
All BTHC-PVC containers allow storage of LR-PCs for up to 9 days with good in vitro parameters. However, the method of sterilization affects the storage conditions of the LR-PCs in these bags.
比较了四种邻苯二甲酸丁酯三己酯增塑聚氯乙烯(BTHC-PVC)容器用于储存白细胞滤除血小板浓缩物(LR-PC)的情况:三种环氧乙烷(EtO)灭菌的(甘布罗、血液技术公司、费森尤斯),以及一种蒸汽灭菌的(费森尤斯)。
LR-PC由5份 Buffy 层和300 ml Composol添加剂溶液制成,并通过过滤进行白细胞滤除。将四个LR-PC合并,随后分到4个BTHC-PVC袋中以防止供体依赖性差异,并在第1、2、5、7和9天取样进行体外分析。
第9天的pH值(平均值±标准差,n = 10)为:7.12±0.03(甘布罗)、7.12±0.04(血液技术公司)和7.07±0.09(费森尤斯,EtO灭菌)(无显著差异),而6.91±0.12(费森尤斯,蒸汽灭菌;与所有EtO灭菌袋相比,P < 0.001)。储存在蒸汽灭菌袋中的LR-PC表现出显著更高的葡萄糖消耗和乳酸生成(与所有EtO灭菌袋相比,P < 0.001)。
所有BTHC-PVC容器都能使LR-PC储存长达9天,且体外参数良好。然而,灭菌方法会影响这些袋子中LR-PC的储存条件。