Gorny K R, Hangiandreou N J, Hesley G K, Felmlee J P
Department of Radiology, 200 First Street SW, Mayo Clinic, Rochester, MN 55905, USA.
Phys Med Biol. 2007 Jan 7;52(1):N13-9. doi: 10.1088/0031-9155/52/1/N02. Epub 2006 Dec 18.
We empirically evaluate mineral oil as an alternative to the mixture of de-gassed water and ultrasound gel, which is currently used as an acoustic coupling medium in clinical magnetic resonance guided focused ultrasound (MRgFUS) treatments. The tests were performed on an ExAblate 2000 MRgFUS system (InSightec Inc., Haifa, Israel) using a clinical patient set-up. Acoustic reflections, treatment temperatures, sonication spot dimensions and position with respect to target location were measured, using both coupling media, in repeated sonications in a tissue mimicking gel phantom. In comparison with the water-gel mix, strengths of acoustic reflections from coupling layers prepared with mineral oil were on average 39% lower and the difference was found to be statistically significant (p = 3.3 x 10(-8)). The treatment temperatures were found to be statistically equivalent for both coupling media, although temperatures corresponding to mineral oil tended to be somewhat higher (on average 1.9 degrees C) and their standard deviations were reduced by about 1 degrees C. Measurements of sonication spot dimensions and positions with respect to target location did not reveal systematic differences. We conclude that mineral oil may be used as an effective non-evaporating acoustic coupling medium for clinical MRgFUS treatments.
我们通过实验评估了矿物油作为脱气水与超声凝胶混合物的替代品的效果,目前在临床磁共振引导聚焦超声(MRgFUS)治疗中,脱气水与超声凝胶的混合物被用作声学耦合介质。测试是在一台ExAblate 2000 MRgFUS系统(以色列海法市InSightec公司)上进行的,采用临床患者设置。在模拟组织的凝胶体模中进行重复超声处理时,使用这两种耦合介质测量了声学反射、治疗温度、超声处理光斑尺寸以及相对于目标位置的光斑位置。与水 - 凝胶混合物相比,用矿物油制备的耦合层的声学反射强度平均低39%,且发现差异具有统计学意义(p = 3.3×10⁻⁸)。尽管与矿物油对应的温度往往略高(平均高1.9摄氏度)且其标准偏差降低了约1摄氏度,但发现两种耦合介质的治疗温度在统计学上相当。关于超声处理光斑尺寸及其相对于目标位置的测量未发现系统性差异。我们得出结论,矿物油可作为临床MRgFUS治疗的一种有效的非蒸发声学耦合介质。