采用预选终椎的脊柱侧弯X线片Cobb角测量比较:传统采集与数字采集

Comparison of Cobb angle measurement of scoliosis radiographs with preselected end vertebrae: traditional versus digital acquisition.

作者信息

Wills Brian P D, Auerbach Joshua D, Zhu Xiaowei, Caird Michelle S, Horn B David, Flynn John M, Drummond Denis S, Dormans John P, Ecker Malcolm L

机构信息

Department of Orthopedics and Rehabilitation, University of Wisconsin, Madison, WI, USA.

出版信息

Spine (Phila Pa 1976). 2007 Jan 1;32(1):98-105. doi: 10.1097/01.brs.0000251086.84420.d1.

Abstract

STUDY DESIGN

Prospective study.

OBJECTIVE

To compare variability in Cobb angle between digitally and traditionally acquired scoliosis radiographs.

SUMMARY OF BACKGROUND DATA

Previous studies have shown that the 95% confidence interval for Cobb angle can vary from 2.6 degrees to 8.8 degrees. No study directly comparing Cobb angles measured from traditional and digitally acquired radiographs has been reported.

METHODS

A spine model simulating 25 single right thoracic curves (range, 20 degrees-60 degrees) was imaged using traditional and digital techniques. Traditional films and miniaturized printed digital films were each measured twice manually. Digital films were also measured twice using computer imaging software.

RESULTS

Overall mean angle and (95% confidence interval) were 41.7 degrees (39.1 degrees-44.3 degrees) for traditional, 40.6 degrees (37.4 degrees-43.8 degrees) for digital, and 39.7 degrees (36.3 degrees-43.1 degrees) for computer measurements. Overall correlation was 0.994 for traditional and digital, 0.987 for traditional and computer, and 0.993 for digital and computer. Fixed effect model analysis demonstrated that, although a statistically significant difference existed between the 3 methods of measuring the Cobb angle (P < 0.0001), the difference between methods was less than 2 degrees.

CONCLUSIONS

Any of the 3 evaluated methods of measurement can be used to measure Cobb angles. Additionally, the methods can be used interchangeably.

摘要

研究设计

前瞻性研究。

目的

比较数字化获取与传统方式获取的脊柱侧弯X光片中Cobb角的变异性。

背景资料总结

以往研究表明,Cobb角的95%置信区间在2.6度至8.8度之间变化。尚无直接比较传统方式与数字化方式获取的X光片中所测Cobb角的研究报道。

方法

使用传统和数字技术对模拟25个单一右胸弯(范围20度至60度)的脊柱模型进行成像。传统胶片和小型打印数字胶片均手动测量两次。数字胶片还使用计算机成像软件测量两次。

结果

传统测量的总体平均角度及(95%置信区间)为41.7度(39.1度至44.3度),数字测量为40.6度(37.4度至43.8度),计算机测量为39.7度(36.3度至43.1度)。传统与数字测量的总体相关性为0.994,传统与计算机测量为0.987,数字与计算机测量为0.993。固定效应模型分析表明,虽然测量Cobb角的3种方法之间存在统计学显著差异(P < 0.0001),但方法之间的差异小于2度。

结论

所评估的3种测量方法中的任何一种均可用于测量Cobb角。此外,这些方法可互换使用。