Zhan Wei, Bard Allen J
Department of Chemistry and Biochemistry, University of Texas, 1 University Station, A5300, Austin, Texas 78712, USA.
Anal Chem. 2007 Jan 15;79(2):459-63. doi: 10.1021/ac061336f.
Liposomes ( approximately 100-nm diameter) containing Ru(bpy)32+ (bpy = 2,2'-bipyridine) were prepared as an electrogenerated chemiluminescent (ECL) tag for a sandwich-type immunoassay of human C-reactive protein (CRP). Polyclonal human CRP antibodies were introduced onto liposomes and magnetic beads through biotin-streptavidin interaction. The antigen-antibody conjugates formed on addition of a CRP-containing sample were separated from unreacted species magnetically. Addition of 0.1 M tri-n-propylamine and 0.1 M phosphate buffer (pH 7.6) containing 0.1 M NaCl and 1% (v/v) Triton X-100 caused liberation of the Ru(bpy)32+ from the liposome. ECL obtained in this medium showed a detection limit of 100 ng/mL for human CRP with good linearity of ECL intensity versus antigen concentration over the range 100 ng/mL-10 microg/mL.
制备了含有Ru(bpy)32+(bpy = 2,2'-联吡啶)的脂质体(直径约100纳米),作为用于人C反应蛋白(CRP)夹心型免疫测定的电化学发光(ECL)标记物。通过生物素-链霉亲和素相互作用将多克隆人CRP抗体引入脂质体和磁珠上。加入含CRP的样品后形成的抗原-抗体结合物通过磁性与未反应的物质分离。加入0.1 M三正丙胺和含有0.1 M NaCl和1%(v/v) Triton X-100的0.1 M磷酸盐缓冲液(pH 7.6)会使Ru(bpy)32+从脂质体中释放出来。在此介质中获得的ECL对人CRP的检测限为100 ng/mL,在100 ng/mL - 10 μg/mL范围内,ECL强度与抗原浓度具有良好的线性关系。