Ochoa Ricardo, Gómez Fredy, Ferreira Ivaldo H, Gutt Federico, de Almeida Carlos E
Laboratório de Ciências Radiológicas (LCR) Universidade do Estado de Rio de Janeiro, Rio de Janeiro, Brazil.
Radiother Oncol. 2007 Feb;82(2):222-8. doi: 10.1016/j.radonc.2007.01.005. Epub 2007 Jan 30.
A new phantom is proposed for measuring the strength of 192Ir high dose rate sources and for verification of the dose calculated by the treatment planning system. The complete formalism and measurement procedure for this phantom is described, as well as the preliminary results obtained in a number of centers around Brazil.
The measurements are performed using powder thermoluminescent dosimeter capsules; the source strength is measured in air and the verification of the dose calculation algorithm in water phantom. The correction factors required to take into account the specificities related to the geometry and the phantom materials have been assessed using the PENELOPE Monte Carlo code and experimental methods. The dedicated phantom, constructed to use as part of a QA program, in this case specifically for high dose rate 192Ir brachytherapy sources, allows simultaneous irradiation of three thermoluminescent dosimeter capsules, requiring only one source stop (dwell positions).
The phantom was mailed to seven radiotherapy institutions in Brazil, and the results show its usefulness in verifying the source air kerma and correctness of treatment planning dose calculation in water phantom.
The comparison made between the phantom measurements, the well-type ionization chamber, and source specifications stated by the hospital (most of the times provided by the source manufacturer) agreed within 3% showing the quality in the HDR dose delivery in Brazilian radiotherapy centers.
提出一种新的模体,用于测量192Ir高剂量率源的强度,并验证治疗计划系统计算的剂量。描述了该模体的完整形式和测量程序,以及在巴西多个中心获得的初步结果。
使用粉末热释光剂量计胶囊进行测量;在空气中测量源强,并在水模体中验证剂量计算算法。使用PENELOPE蒙特卡罗代码和实验方法评估了考虑与几何形状和模体材料相关的特异性所需的校正因子。专门构建的模体用作质量保证计划的一部分,在这种情况下专门用于高剂量率192Ir近距离放射治疗源,允许同时照射三个热释光剂量计胶囊,只需要一个源停止(驻留位置)。
该模体被邮寄到巴西的七个放射治疗机构,结果表明其在验证源空气比释动能和水模体中治疗计划剂量计算的正确性方面很有用。
模体测量、井型电离室与医院(大多数情况下由源制造商提供)规定的源规格之间的比较在3%以内相符,表明巴西放射治疗中心高剂量率剂量传递的质量。