Measurement of the leakage radiation from linear accelerators in the backward direction for 4, 6, 10, 15, and 18 MV x-ray energies.

作者信息

Jaradat Adnan K, Biggs Peter J

机构信息

Department of Physics, University of Massachusetts Lowell, One University Avenue, Lowell, MA 01854, USA.

出版信息

Health Phys. 2007 Apr;92(4):387-95. doi: 10.1097/01.HP.0000252308.89839.7d.

Abstract

The x-ray leakage from the housing of a therapy x-ray source is regulated to be <0.1% of the useful beam exposure at 1 m from the source. It is to be expected that the machine leakage in the backward direction would be less because the gantry and stand contain significant amounts of additional metal to attenuate the x rays. A reduction in head leakage in this direction will have a direct effect on the thickness of the shielding wall behind the linear accelerator. However, no reports have been published to date on measurements in this area. The x-ray leakage in the backward direction has been measured from linacs having energies of 4, 6, 10, 15, and 18 MV using a 100 cm ionization chamber and Al2O3 dosimeters. The leakage was measured at nine different positions over the rear wall using a 3 x 3 matrix with a 1-m separation between adjacent horizontal and vertical points with either the leftmost or rightmost column aligned with the target and isocenter. In general, the leakage is less than the canonical value, but the exact value depends on energy, gantry angle, and measurement position. There is significantly greater attenuation directly behind the gantry stand for all energies. Leakage at 10 MV for some positions exceeded 0.1%. Additionally, neutron leakage measurements were made for 10, 15, and 18 MV x-ray beams using track-etch detectors. The average neutron leakage was less than 0.1% except for 18 MV, where neutron leakage was more than 0.1% of the useful beam at some positions.

摘要

治疗用X射线源外壳的X射线泄漏量被规定为在距离源1米处小于有用线束照射量的0.1%。由于机架和支架含有大量额外的金属来衰减X射线,预计机器在向后方向的泄漏会更少。该方向头部泄漏的减少将直接影响直线加速器后方屏蔽墙的厚度。然而,迄今为止尚未有关于该区域测量的报告发表。已使用100厘米电离室和Al2O3剂量计对能量为4、6、10、15和18兆伏的直线加速器在向后方向的X射线泄漏进行了测量。使用3×3矩阵在整个后壁的九个不同位置进行泄漏测量,相邻水平和垂直点之间的间距为1米,最左边或最右边的列与靶和等中心对齐。一般来说,泄漏小于规定值,但具体数值取决于能量、机架角度和测量位置。对于所有能量,在机架支架正后方的衰减明显更大。在某些位置,10兆伏时的泄漏超过了0.1%。此外,还使用径迹蚀刻探测器对10、15和18兆伏X射线束的中子泄漏进行了测量。除了18兆伏时在某些位置中子泄漏超过有用线束的0.1%外,平均中子泄漏小于0.1%。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验