Donard O F X, Bruneau F, Moldovan M, Garraud H, Epov V N, Boust D
Laboratoire de Chimie Bio-Inorganique et Environnement, CNRS UMR 5034, Université de Pau et des Pays de l'Adour, Helioparc Pau Pyrénées, 2 avenue du Président Angot, 64053 Pau Cedex 9, France.
Anal Chim Acta. 2007 Mar 28;587(2):170-9. doi: 10.1016/j.aca.2007.01.047. Epub 2007 Jan 25.
Among the transuranic elements present in the environment, plutonium isotopes are mainly attached to particles, and therefore they present a great interest for the study and modelling of particle transport in the marine environment. Except in the close vicinity of industrial sources, plutonium concentration in marine sediments is very low (from 10(-4) ng kg(-1) for (241)Pu to 10 ng kg(-1) for (239)Pu), and therefore the measurement of (238)Pu, (239)Pu, (240)Pu, (241)Pu and (242)Pu in sediments at such concentration level requires the use of very sensitive techniques. Moreover, sediment matrix contains huge amounts of mineral species, uranium and organic substances that must be removed before the determination of plutonium isotopes. Hence, an efficient sample preparation step is necessary prior to analysis. Within this work, a chemical procedure for the extraction, purification and pre-concentration of plutonium from marine sediments prior to sector-field inductively coupled plasma mass spectrometry (SF-ICP-MS) analysis has been optimized. The analytical method developed yields a pre-concentrated solution of plutonium from which (238)U and (241)Am have been removed, and which is suitable for the direct and simultaneous measurement of (239)Pu, (240)Pu, (241)Pu and (242)Pu by SF-ICP-MS.
在环境中存在的超铀元素中,钚同位素主要附着于颗粒上,因此对于海洋环境中颗粒输运的研究和建模而言,它们具有极大的研究价值。除了在工业源附近区域外,海洋沉积物中的钚浓度非常低((241)Pu为10(-4) ng kg(-1),(239)Pu为10 ng kg(-1)),因此要在这样的浓度水平下测量沉积物中的(238)Pu、(239)Pu、(240)Pu、(241)Pu和(242)Pu,需要使用非常灵敏的技术。此外,沉积物基质含有大量的矿物种类、铀和有机物质,在测定钚同位素之前必须将其去除。因此,在分析之前需要一个高效的样品制备步骤。在这项工作中,对一种用于在扇形磁场电感耦合等离子体质谱(SF-ICP-MS)分析之前从海洋沉积物中提取、纯化和预浓缩钚的化学程序进行了优化。所开发的分析方法得到了一种预浓缩的钚溶液,其中已去除了(238)U和(241)Am,并且适合通过SF-ICP-MS直接同时测量(239)Pu、(240)Pu、(241)Pu和(242)Pu。