Electrode comparison in pattern electroretinography.

作者信息

Prager T C, Saad N, Schweitzer F C, Garcia C A, Arden G B

机构信息

Department of Ophthalmology, University of Texas Medical School, Houston.

出版信息

Invest Ophthalmol Vis Sci. 1992 Feb;33(2):390-4.

PMID:1740370
Abstract

In recent years, there has been great interest in recording the pattern electroretinogram (PERG) in glaucomatous and diabetic populations. The Dawson, Trick, and Litzkow thread electrode (DTLTE) and the gold foil electrode (GFE), commonly used for recording PERGs, were compared for variations in amplitude of response, test-retest variability, and patient comfort. Two study centers collected data on a total of 32 normal subjects. The subjects from the London center showed a slight (but not significant) preference for the DTLTE, and the Houston subjects also found the DTLTE to be significantly more comfortable (chi-square = 39, P less than 0.001). In both study groups, the GFE was found to produce a statistically larger amplitude of response than that obtained with the DTLTE. Significant differences were found regardless of the slow (transient, 3.1 Hz; F = 6.24; P = 0.0192) or fast (steady state, 8.3 Hz; F = 18.38; P = 0.0001) stimulus-presentation rate. Larger differences between the two electrodes occurred under steady-state conditions. Although there is no consensus as to the optimum recording conditions to obtain the subtle PERG, it appears the the GFE records larger responses than the DTLTE. However, test-retest data confirmed that the GFE records twice the amplitude of the DTLTE, and it also produced twice the variability (average percent difference over time for GFE, 15%; for DTLTE, 8%).

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验