Tokumine Joho, Nitta Kenichi, Higa Tatsuya, Gushiken Kouji, Okayama Haruka, Sugahara Kazuhiro
Department of Anesthesia, Ryukyu University Hospital, Okinawa.
Masui. 2007 Apr;56(4):453-8.
We checked for the occurrence of any leakage in an anesthesia circuit to estimate the present situation of airtightness of an anesthesia circuit in 55 anesthesia machines at various medical institutions in Okinawa.
Leakage was detected and measured by a leak checker.
A certain degree of leakage was found in 28 of the anesthesia machines (51%). The average and maximal amounts of leakage in the anesthesia machines were 0.26, and 1.29 l x min(-1), respectively. Thirty-eight of anesthesia machines met the standards of a low flow leak test (69%). The average and maximal amounts of leakage in the inside circuit were 0.02, and 0.20 l x min(-1), respectively. The leak points of the inside circuit were present in a unidirectional valve for an oxygen sensor (14%) and for unknown reason (7%). The average and maximal amounts of leakage in the outside circuit were 0.24, and 1.29 l x min(-1), respectively. Leak points in the inside circuit were found in the corrugated breathing tube (82%) and canister (32%).
Most instances of leakage of the above described anesthesia circuits were identified in the corrugated breathing tube and canister. As a result of our findings, we strongly recommend that these parts be checked very carefully during pre-anesthetic leak tests.
我们检查了麻醉回路中是否存在泄漏情况,以评估冲绳县各医疗机构55台麻醉机的麻醉回路气密性现状。
使用检漏仪检测并测量泄漏情况。
在28台麻醉机(51%)中发现了一定程度的泄漏。麻醉机的平均泄漏量和最大泄漏量分别为0.26和1.29升/分钟。38台麻醉机符合低流量泄漏测试标准(69%)。内部回路的平均泄漏量和最大泄漏量分别为0.02和0.20升/分钟。内部回路的泄漏点出现在氧气传感器单向阀处(14%)以及原因不明处(7%)。外部回路的平均泄漏量和最大泄漏量分别为0.24和1.29升/分钟。内部回路的泄漏点出现在波纹呼吸管(82%)和滤罐(32%)。
上述麻醉回路的泄漏情况大多出现在波纹呼吸管和滤罐中。根据我们的研究结果,我们强烈建议在麻醉前泄漏测试期间仔细检查这些部件。