[Hypoventilation due to disconnection of the vaporizer and negative-pressure leak test to find disconnection].

作者信息

Yasukawa M, Yasukawa K

机构信息

Department of Anesthesia, Iwamizawa Municipal General Hospital.

出版信息

Masui. 1992 Aug;41(8):1345-9.

PMID:1433865
Abstract

Most of Ohmeda anesthesia machines have a outlet check valve located in the low-pressure circuit, and this influences the type of leak test greatly. We report a case of hypoventilation due to a gas leakage from the junction of two vaporizers, Fluotec 3 and Enfluratec of Ohio 30/70. This gas leakage was not discovered preoperatively by performing traditional leak test of squeezing the bag filled with oxygen. Following intubation of the trachea the patient was paralyzed and mechanically ventilated. Her vital signs were stable and operation was started. Soon thereafter, she became hypertensive in spite of increases in concentration of enflurane and supplemental doses of fentanyl and diltiazem. ETCO2 showed 7.4% and spontaneous respiration started. We checked anesthesia machine and found a gas leakage at the junction of two vaporizers. Anesthesia machine was exchanged. The remainder of the procedure was uneventful and the patient showed a normal recovery. The outlet check valve closes when back-pressure is exerted on it, and a vulnerable area including flow tubes and vaporizers is not tested by the traditional positive-pressure leak test squeezing bag. Ohmeda recommends the use of a negative pressure leak test with suction bulb device. We propose manufacturers to place a cation on anesthesia machine such as -- "Negative-pressure leak test is inevitable because of presence of outlet check valve."

摘要

相似文献

1
[Hypoventilation due to disconnection of the vaporizer and negative-pressure leak test to find disconnection].
Masui. 1992 Aug;41(8):1345-9.
2
[Leak test for the internal circuit of anesthesia machines].[麻醉机内部回路泄漏测试]
Masui. 1999 May;48(5):542-7.
3
[Leak of anesthetic gases inside the anesthesia machine].[麻醉机内麻醉气体泄漏]
Masui. 2006 May;55(5):623-5.
4
Assessment of vaporizer performance in low-flow and closed-circuit anesthesia.低流量和闭环麻醉中蒸发器性能的评估。
Anesth Analg. 1980 May;59(5):359-66.
5
A comparison of the accuracy of the Siemens injection vaporizer and the Enfluratec 3 vaporizer.
Anaesthesiol Reanim. 1994;19(3):75-7.
6
[Gas leakage from an anesthetic circuit caused by mis-installation of a canister].[因罐安装错误导致麻醉回路气体泄漏]
Masui. 2008 Nov;57(11):1427-30.
7
[An unusual cause of anesthetic ventilator malfunction].[麻醉呼吸机故障的一个不寻常原因]
Masui. 1992 Dec;41(12):1994-7.
8
[Where is a leak point detected by "the low flow leak test" of anesthetic machines?].
Masui. 1997 Oct;46(10):1385-9.
9
[A new leak test of anesthetic machine].
Masui. 1993 Sep;42(9):1375-7.
10
[Anesthetic machine leakage from vaporizer by external force derived from keyboard of electronic medical records].
Masui. 2012 Nov;61(11):1288-90.