Suppr超能文献

测量和监测早产儿心电图带张力以评估呼吸功能。

Measurement and monitoring of electrocardiogram belt tension in premature infants for assessment of respiratory function.

作者信息

Ciaccio Edward J, Hiatt Mark, Hegyi Thomas, Drzewiecki Gary M

机构信息

Department of Pharmacology, Columbia University, New York, USA.

出版信息

Biomed Eng Online. 2007 Apr 19;6:13. doi: 10.1186/1475-925X-6-13.

Abstract

BACKGROUND

Monitoring of the electrocardiogram (ECG) in premature infants with conventional adhesive-backed electrodes can harm their sensitive skin. Use of an electrode belt prevents skin irritation, but the effect of belt pressure on respiratory function is unknown. A strain gauge sensor is described which measures applied belt tension.

METHOD

The device frame was comprised of an aluminum housing and slide to minimize the device weight. Velcro tabs connected housing and slide to opposite tabs located at the electrode belt ends. The slide was connected to a leaf spring, to which were bonded two piezoresistive transducers in a half-bridge circuit configuration. The device was tested for linearity and calibrated. The effect on infant respiratory function of constant belt tension in the normal range (30 g-90 g) was determined.

RESULTS

The mechanical response to a step input was second order (fn = 401 Hz, zeta = 0.08). The relationship between applied tension and output voltage was linear in the range 25-225 gm of applied tension (r2 = 0.99). Measured device sensitivity was 2.18 mV/gm tension using a 5 V bridge excitation voltage. When belt tension was increased in the normal range from 30 gm to 90 gm, there was no significant change in heart rate and most respiratory functions during monitoring. At an intermediate level of tension of 50 gm, pulmonary resistance and work of breathing significantly decreased.

CONCLUSION

The mechanical and electrical design of a device for monitoring electrocardiogram electrode belt tension is described. Within the typical range of application tension, cardiovascular and respiratory function are not substantially negatively affected by electrode belt force.

摘要

背景

使用传统的背胶电极监测早产儿的心电图可能会损伤其敏感皮肤。使用电极带可防止皮肤刺激,但电极带压力对呼吸功能的影响尚不清楚。本文描述了一种用于测量电极带施加张力的应变片传感器。

方法

该装置框架由铝制外壳和滑块组成,以尽量减轻装置重量。尼龙搭扣将外壳和滑块连接到位于电极带两端的相对搭扣上。滑块连接到一个板簧上,在该板簧上以半桥电路配置粘贴了两个压阻式传感器。对该装置进行线性测试并校准。确定了正常范围内(30克至90克)恒定电极带张力对婴儿呼吸功能的影响。

结果

对阶跃输入的机械响应为二阶响应(固有频率fn = 401赫兹,阻尼比zeta = 0.08)。在25至225克的施加张力范围内,施加张力与输出电压之间的关系呈线性(r2 = 0.99)。使用5伏桥式激励电压时,测得的装置灵敏度为2.18毫伏/克张力。当电极带张力在正常范围内从30克增加到90克时,监测期间心率和大多数呼吸功能无显著变化。在50克的中间张力水平下,肺阻力和呼吸功显著降低。

结论

本文描述了一种用于监测心电图电极带张力的装置的机械和电气设计。在典型的应用张力范围内,电极带的压力对心血管和呼吸功能没有实质性的负面影响。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/1312/1868740/e285400ac94e/1475-925X-6-13-1.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验