Hennekinne J-A, Guillier F, Perelle S, De Buyser M-L, Dragacci S, Krys S, Lombard B
Laboratory for Study and Research on Quality of Food and on Food Processes, French Agency for Food Safety (AFSSA-lerqap), National and Community Reference Laboratory, Maisons-Alfort cedex, France.
J Appl Microbiol. 2007 May;102(5):1261-72. doi: 10.1111/j.1365-2672.2006.03183.x.
Immunological tools used to detect staphylococcal enterotoxins (SEs) in foods are numerous. The aim of this study was to evaluate, on naturally contaminated milk product samples, the performance of the Vidas SET2, in comparison to the Transia plate SET.
The Vidas SET2 was compared with the Transia plate SET on supernatants of Staphylococcus aureus isolates and on naturally contaminated milk products. It is noteworthy that when using IgG rabbit treatment, both kits can be considered as equivalent to detect enterotoxins in naturally contaminated milk products.
This study demonstrated that the Vidas SET2 performance is similar to that of Transia plate SET kit, when a rabbit IgG treatment step is used before detection step. This additional treatment significantly decreased, from 42% to 8%, the rate of positive deviations observed using the Transia plate SET detection kit.
The Vidas SET2 was clearly found as more specific, when no preliminary rabbit IgG treatment was used, and which results in a better workflow when a large number of samples have to be analysed within a few days. Considering the results obtained, the Vidas SET2 detection kit can be used to assess the safety of milk products for SEs.
用于检测食品中葡萄球菌肠毒素(SEs)的免疫学工具众多。本研究的目的是在天然污染的奶制品样本上,评估Vidas SET2相较于Transia plate SET的性能。
在金黄色葡萄球菌分离株的上清液以及天然污染的奶制品上,对Vidas SET2和Transia plate SET进行了比较。值得注意的是,当使用兔IgG处理时,两种试剂盒在检测天然污染奶制品中的肠毒素方面可被视为等效。
本研究表明,在检测步骤前使用兔IgG处理步骤时,Vidas SET2的性能与Transia plate SET试剂盒相似。这一额外处理显著降低了使用Transia plate SET检测试剂盒时观察到的正偏差率,从42%降至8%。
当不使用兔IgG预处理时,Vidas SET2显然更具特异性,并且当必须在几天内分析大量样本时,其工作流程更佳。考虑到所获得的结果,Vidas SET2检测试剂盒可用于评估奶制品中SEs的安全性。