Low-dose protocol for head CT in monitoring hydrocephalus in children.

作者信息

Rybka Krzysztof, Staniszewska Anna Maria, Biegański Tadeusz

机构信息

Department of Diagnostic Imaging, Polish Mother's Memorial Hospital - Research Institute, Łódź, Poland.

出版信息

Med Sci Monit. 2007 May;13 Suppl 1:147-51.

Abstract

BACKGROUND

Children with hydrocephalus and increased intracranial pressure are treated with ventricular shunts. Multiple head CT is used to assess shunt malfunction and changes in ventricular size (volume). Because of the naturally high contrast of the head structures in CT imaging, a low-dose protocol for hydrocephalic pediatric patients was proposed. The aim of the study was to assess the implemented protocol in categories of radiation risk to the patients.

MATERIAL/METHODS: Head CT examinations were performed using the fourth-generation single-slice scanner PQ-2000 (Picker). Exposure parameters of the routine head procedure stored in the scanner software were the starting point for the investigation. The acceptable ranges of the modified parameters were found on the basis of earlier published results. The effect of the new protocol on image quality was tested using a CATPHAN phantom. The organ doses were evaluated experimentally. The clinical images were evaluated by two experienced and independent radiologists.

RESULTS

Head CT examinations were performed in 380 children with hydrocephalus during a 10-year period at the Department of Diagnostic Imaging Polish Mother's Memorial Hospital. Until 2005 these children were examined according to the routine protocol of the scanner, i.e. 10 mm contiguous axial, 120 kVp, and 225 mAs. Then the protocol was modified to 10 mm contiguous axial, 100 kVp, and 150 mAs. Reductions in patient doses to 70% were found while the clinical images maintained sufficient diagnostic value.

CONCLUSIONS

The low-dose protocol for head CT can be recommended for the monitoring of pediatric hydrocephalus patients.

摘要

背景

脑积水及颅内压升高的儿童采用脑室分流术进行治疗。多次头颅CT用于评估分流器故障及脑室大小(容积)的变化。由于CT成像中头部结构天然具有高对比度,因此提出了针对脑积水患儿的低剂量扫描方案。本研究的目的是评估所实施的方案对患者辐射风险的影响。

材料/方法:使用第四代单排扫描仪PQ - 2000(Picker)进行头颅CT检查。以存储在扫描仪软件中的常规头部检查的曝光参数作为研究起点。根据先前发表的结果确定修改后参数的可接受范围。使用CATPHAN体模测试新方案对图像质量的影响。通过实验评估器官剂量。由两名经验丰富且独立的放射科医生对临床图像进行评估。

结果

在波兰母亲纪念医院诊断影像科的10年期间,对380名脑积水儿童进行了头颅CT检查。直到2005年,这些儿童按照扫描仪的常规方案进行检查,即10mm连续轴位扫描、120kVp和225mAs。然后将方案修改为10mm连续轴位扫描、100kVp和150mAs。发现患者剂量降低了70%,而临床图像仍保持足够的诊断价值。

结论

头颅CT低剂量扫描方案可推荐用于小儿脑积水患者的监测。