Sasaki T, Takatani S, Shiono M, Sakuma I, Noon G P, Nosé Y, DeBakey M E
Department of Surgery, Baylor College of Medicine, Houston, Texas 77030.
ASAIO Trans. 1991 Jul-Sep;37(3):M249-51.
An implantable electromechanical ventricular assist system (VAS) intended for permanent human use was developed. It consisted of a conically shaped pumping chamber, a polyolefin (Hexsyn) rubber diaphragm attached to a pusher-plate, and a compact actuator with a direct current brushless motor and a planetary rollerscrew. The outer diameter was 97 mm, and the total thickness was 70 mm. This design was chosen to give a stroke volume of 63 ml. The device weighs 620 g, with a total volume of 360 ml. The pump can provide 8 L/min flow against 120 mmHg afterload with a preload of 10 mmHg. The inner surface of the device, including the pumping chamber and diaphragm, was made biocompatible with a dry gelatin coating. To date, two subacute (2 and 6 day) calf studies have been conducted. The pump showed reasonable anatomic fit inside the left thorax, and the entire system functioned satisfactorily in both the fill-empty mode using the Hall effect sensor signals and the conventional fixed rate mode. There were no thromboembolic complications despite no anticoagulation therapy. The system now is being endurance tested greater than 10 weeks (9 million cycles). This VAS is compact, low noise, easy to control, and has excellent biocompatibility.
开发了一种旨在供人类长期使用的植入式机电心室辅助系统(VAS)。它由一个圆锥形泵腔、一个连接到推板的聚烯烃(Hexsyn)橡胶隔膜以及一个带有直流无刷电机和行星滚柱丝杠的紧凑型致动器组成。外径为97毫米,总厚度为70毫米。选择这种设计是为了使 stroke volume达到63毫升。该装置重620克,总体积为360毫升。该泵在10毫米汞柱的前负荷下,可在120毫米汞柱后负荷下提供8升/分钟的流量。该装置的内表面,包括泵腔和隔膜,通过干燥明胶涂层实现了生物相容性。迄今为止,已经进行了两项亚急性(2天和6天)小牛研究。该泵在左胸内显示出合理的解剖适应性,并且整个系统在使用霍尔效应传感器信号的填充-排空模式和传统固定速率模式下均运行良好。尽管没有进行抗凝治疗,但没有血栓栓塞并发症。该系统目前正在进行超过10周(900万次循环)的耐久性测试。这种VAS紧凑、低噪音、易于控制,并且具有出色的生物相容性。