Cyr Steven J, Currier Bradford L, Eck Jason C, Foy Andrew, Chen Qingshan, Larson Dirk R, Yaszemski Michael J, An Kai-Nan
Department of Orthopedic Surgery, Mayo Clinic College of Medicine, 200 First Street SW, Rochester, MN 55905, USA.
Spine J. 2008 Jul-Aug;8(4):661-5. doi: 10.1016/j.spinee.2007.02.008. Epub 2007 May 11.
The internal carotid artery and hypoglossal nerve lie over the anterior aspect of the lateral mass of the atlas and are at risk from bicortical C1-C2 transarticular screws. This has led to the recommendation for unicortical screws if the neurovascular structures are in close proximity to the proposed exit point. No data are available on strength of unicortical versus bicortical C1-C2 transarticular screws.
To compare the biomechanical pullout strength of unicortical versus bicortical C1-C2 transarticular screws in a cadaveric model.
Biomechanical study.
Fifteen cervical spine specimens underwent axial pullout testing. A unicortical C1-C2 transarticular screw was placed on one side with a contralateral bicortical screw. Data were analyzed to reveal any significant differences in strength.
Mean pullout strength for the bicortical C1-C2 transarticular screws was 1,048.8 (+/-360.1) N versus 939.2 (+/-360.6) for unicortical screws (p=.22). There was no significant difference in the pullout strength of unicortical and bicortical screws.
In cases with satisfactory bone quality, it appears reasonable to use unicortical screws to avoid the risk of neurovascular injury from penetrating the anterior cortex of C1.
颈内动脉和舌下神经位于寰椎侧块的前方,存在双皮质C1-C2经关节螺钉置入时的风险。如果神经血管结构靠近拟置入点,这就导致了推荐使用单皮质螺钉。目前尚无关于单皮质与双皮质C1-C2经关节螺钉强度的数据。
在尸体模型中比较单皮质与双皮质C1-C2经关节螺钉的生物力学拔出强度。
生物力学研究。
对15个颈椎标本进行轴向拔出试验。一侧置入单皮质C1-C2经关节螺钉,对侧置入双皮质螺钉。分析数据以揭示强度上的任何显著差异。
双皮质C1-C2经关节螺钉的平均拔出强度为1048.8(±360.1)N,而单皮质螺钉为939.2(±360.6)N(p = 0.22)。单皮质和双皮质螺钉的拔出强度无显著差异。
在骨质质量良好的情况下,使用单皮质螺钉以避免穿透C1前皮质造成神经血管损伤似乎是合理的。