Comparison of dose calibration by ionometric and chemical dosimetry for 6- and 45-MV x-ray beams.

作者信息

Zwicker R D, Wu A, Sternick E S

机构信息

Medical College of Virginia, Richmond 23298.

出版信息

Med Phys. 1991 Nov-Dec;18(6):1251-3. doi: 10.1118/1.596599.

Abstract

This work reports the results of an experiment carried out to investigate the compatibility of TG-21 ion chamber calibration with Fricke dosimetry near the extremes of the clinically available megavoltage photon energy range. Doses determined from ion chamber measurements and from chemical dosimetry are compared for a 6-MV linear accelerator x-ray beam and 45-MV betatron beam. The Fricke dosimetry was carried out by the National Institute of Standards and Technology. Agreement within 1% between the two techniques was obtained at both photon energies using a currently accepted G-value.

摘要

本研究报告了一项实验的结果,该实验旨在研究在临床可用的兆伏级光子能量范围的极值附近,TG-21电离室校准与弗里克剂量测定法的兼容性。针对6兆伏直线加速器X射线束和45兆伏电子感应加速器束,比较了通过电离室测量和化学剂量测定法确定的剂量。弗里克剂量测定法由美国国家标准与技术研究院进行。在两种光子能量下,使用当前公认的G值,两种技术之间的一致性在1%以内。