Zwicker R D, Wu A, Sternick E S
Medical College of Virginia, Richmond 23298.
Med Phys. 1991 Nov-Dec;18(6):1251-3. doi: 10.1118/1.596599.
This work reports the results of an experiment carried out to investigate the compatibility of TG-21 ion chamber calibration with Fricke dosimetry near the extremes of the clinically available megavoltage photon energy range. Doses determined from ion chamber measurements and from chemical dosimetry are compared for a 6-MV linear accelerator x-ray beam and 45-MV betatron beam. The Fricke dosimetry was carried out by the National Institute of Standards and Technology. Agreement within 1% between the two techniques was obtained at both photon energies using a currently accepted G-value.