Rohlfing Andreas, Müthing Johannes, Pohlentz Gottfried, Distler Ute, Peter-Katalinić Jasna, Berkenkamp Stefan, Dreisewerd Klaus
Institute of Medical Physics and Biophysics, Westfälische Wilhelms-Universität Münster, Robert-Koch-Strasse 31, 48149 Münster, Germany.
Anal Chem. 2007 Aug 1;79(15):5793-808. doi: 10.1021/ac070633x. Epub 2007 Jun 23.
The application of a recently developed direct coupling of high-performance thin-layer chromatography (HPTLC) and infrared matrix-assisted laser desorption/ionization orthogonal extracting time-of-flight mass spectrometry (Dreisewerd, K.; Müthing, J.; Rohlfing, A.; Meisen, I.; Vukelic, Z.; Peter-Katalinic, J.; Hillenkamp, F.; Berkenkamp, S. Anal. Chem. 2005, 77, 4098-4107) to the analysis of phospholipid mixtures is demonstrated. Mixtures of six phospholipid types were exemplarily analyzed. The sensitivity was found to be in the range between about 10 and 150 pmol of material spotted for HPTLC, depending on phospholipid acidity, Rf value, and ion polarity. The lateral resolution of the analysis is on the order of the laser focus diameter of about 220 x 300 microm2, allowing differentiation between phospholipid species of different acyl chain composition within one single HPTLC band, which were undistiguishable by a mere visual assessment. Analyte diffusion due to the addition of glycerol to the HPTLC plate was found to be-if at all notable-of only minor importance.
本文展示了一种最近开发的将高效薄层色谱法(HPTLC)与红外基质辅助激光解吸/电离正交提取飞行时间质谱法直接联用的技术(德赖泽韦德,K.;米辛,J.;罗尔芬,A.;梅森,I.;武凯利奇,Z.;彼得 - 卡塔林奇,J.;希伦坎普,F.;贝肯坎普,S.《分析化学》2005年,77卷,4098 - 4107页)在磷脂混合物分析中的应用。以六种磷脂类型的混合物为例进行了分析。结果发现,根据磷脂酸度、比移值(Rf)和离子极性,HPTLC点样物质的灵敏度在约10至150皮摩尔之间。分析的横向分辨率约为激光聚焦直径220×300平方微米,这使得在单个HPTLC条带内能够区分不同酰基链组成的磷脂种类,而仅通过目视评估是无法区分的。发现向HPTLC板中添加甘油导致的分析物扩散(如果确实显著的话),其影响微不足道。