A 16-channel 8-parameter waveform electrotactile stimulation system.

作者信息

Kaczmarek K A, Kramer K M, Webster J G, Radwin R G

机构信息

Department of Electrical and Computer Engineering, University of Wisconsin, Madison 53706.

出版信息

IEEE Trans Biomed Eng. 1991 Oct;38(10):933-43. doi: 10.1109/10.88439.

Abstract

We have developed a general-purpose electrotactile (electrocutaneous) stimulation system as a research tool for studying psychophysiological performance associated with various stimulation waveforms. An experimenter-defined command file specifies the stimulation current and waveform of each of the 16 channels. The system provides burst onset delay of 0-20 ms, phase current of 0-50 mA, interphase interval of 0-1000 microseconds, number of pulses per burst from 1-100, pulse repetition rate of 0.1-25 kHz, phase width of 2-1000 microseconds, and functionally-monophasic pulses (with zero dc current) or balanced-biphasic pulses (with equal positive and negative phases). The system automatically delivers the desired stimulation, prompts the subject for responses, and then logs subject responses. Key features of the system are 1) very flexible choice of bursts of pulsatile waveforms, 2) real-time control of all of the waveform parameters as mathematical functions of external analog inputs, and 3) high-performance electrode-driver circuitry.