Suppr超能文献

A four-channel IBM PC/AT compatible biphasic pulse generator for nerve stimulation.

作者信息

Meadows P M, McNeal D R

出版信息

IEEE Trans Biomed Eng. 1989 Jul;36(7):802-4. doi: 10.1109/10.32115.

Abstract

An IBM PC/AT compatible four-channel biphasic pulse generator has been developed to assist in functional electrical stimulation (FES) related research. Each channel uses a bipolar 12-bit digital-to-analog converter (DAC) to generate biphasic voltage pulses between +/- 5 VDC and three 16-bit timers to control first, second, and interphase durations. Two vectored interrupt generators are available for precise pulse timing control. Potential uses of this device include FES research, the characterization of the recruitment properties of percutaneous electrodes in multichannel stimulation systems, or as a subassembly in commercial medical devices requiring electrical simulation in an IBM PC/AT platform.