Osborne R W, Durkin T A
Department of Veterinary Pathology, Glasgow University Veterinary School, Rossendale, Lancs, UK.
J Appl Bacteriol. 1991 Nov;71(5):434-8. doi: 10.1111/j.1365-2672.1991.tb03813.x.
The considerable refinements necessary to enable Class I and II microbiological safety cabinets to operate in a force-ventilated laboratory and to meet appropriate safety criteria have been reported previously. The continued successful operation of such cabinets without a deterioration of operator protection is described. The performance of two Class II units, one meeting and one failing the current British Standard applied to four head KI-discus testing, is compared and discussed. In addition, some further potential difficulties within the environment, which could compromise cabinet containment, are highlighted.